62 MAX/62 MAX + Infrared Thermometer Users Manual PN 4060712 April 2012 © 2012 Fluke Corporation. All rights reserved. Printed in China. Specifications are subject to change without notice. All product names are trademarks of their respective companies.
LIMITED WARRANTY AND LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke’s behalf.
Table of Contents Title Page Introduction ................................................................................................................. 1 How to Contact Fluke ................................................................................................. 1 Safety Information ...................................................................................................... 2 Maintenance ...............................................................................................
62 MAX/62 MAX + Users Manual ii
Introduction The Fluke 62 MAX and 62 MAX + Infrared Thermometers (the Product) can determine the surface temperature by measuring the amount of infrared energy radiated by the target’s surface. Note that the Japanese models indicate Celsius only. WWarning Read all safety information before you use the Product.
2 MAX/62 MAX + Users Manual Or, visit Fluke's website at www.fluke.com. To register your product, visit http://register.fluke.com. To see, print, or download the latest manual supplement, visit http://us.fluke.com/usen/support/manuals. Safety Information A Warning identifies conditions and procedures that are dangerous to the user. A Caution identifies conditions and procedures that can cause damage to the Product or the equipment under test.
Infrared Thermometer Safety Information • • • • • • See emissivity information for actual temperatures. Reflective objects result in lower than actual temperature measurements. These objects pose a burn hazard. Do not look directly into the laser with optical tools (for example, binoculars, telescopes, microscopes). Optical tools can focus the laser and be dangerous to the eye. Do not look into the laser. Do not point laser directly at persons or animals or indirectly off reflective surfaces.
62 MAX/62 MAX + Users Manual Table 1. Symbols Symbol Symbol Meaning W Risk of danger. Important information. See Manual. ~ Do not dispose of this product as unsorted municipal waste. Go to Fluke’s website for recycling information. * Warning. Laser. P Conforms to European Union directives. Battery Conforms to relevant Australian standards.
Infrared Thermometer Maintenance Maintenance WCaution To avoid damage to the Product, do not leave the thermometer on or near objects of high temperature. How to Change the Battery To install or change the AA IEC LR06 battery, open the battery compartment and replace the battery as shown in Figure 16. How to Clean the Product Use soap and water on a damp sponge or soft cloth to clean the Product case. Carefully wipe the surface with a moist cotton swab. The swab may be moistened with water. See Figure 17.
62 MAX/62 MAX + Users Manual Specifications Temperature Range Accuracy Response Time (95 %) Spectral Response Emissivity 6 62 MAX 62 MAX + -30 °C to 500 °C (-22 °F to 932 °F) -30 °C to 650 °C (-22 °F to 1202 °F) ≥0 °C: ±1.5 °C or ±1.5 % of reading, whichever is greater (≥32 °F: ±3 °F or ±1.5 % of reading, whichever is greater) ≥ -10 °C to <0 °C: ±2 °C (≥14 °F to <32 °F: ±4 °F) < -10 °C: ±3 °C (<14 °F: ±6 °F) <500 ms (95 % of reading) 8 to 14 microns 0.10 to 1.
Infrared Thermometer Specifications Optical Resolution 10:1 (calculated at 90 % energy) Display Resolution 0.1 °C (0.2 °F) 12:1 (calculated at 90 % energy) Weight Size ±0.8 % of reading or ±1.0 °C ±0.5 % of reading or ±0.5 °C (2 °F), whichever is greater (1 °F), whichever is greater 1 AA IEC LR06 Battery 10 hours with laser and 8 hours with laser and backlight on backlight on 255 g (8.99 oz) (175 x 85 x 75) mm (6.88 x 3.34 x 2.
62 MAX/62 MAX + Users Manual IP Rating Drop Test Vibration and Shock EMC IP 54 per IEC 60529 3 meters IEC 68-2-6 2.5 g, 10 to 200 Hz, IEC 68-2-27, 50 g, 11 ms EN 61326-1:2006 EN 61326-2:2006 Standards and Agency Approval Compliance ............................................. EN/IEC 61010-1: 2001 Laser Safety ............................................
Infrared Thermometer The Product The Product 1 2 gtt01.
62 MAX/62 MAX + Users Manual 3 4 / 5 5 6 Max/Min/Avg/Diff gtt14.
Infrared Thermometer The Product 7 + = 10 min OFF gtt16.
62 MAX/62 MAX + Users Manual 8 10 0ºC 32ºF 9 30 Min 11 12 9 3 6 22ºC 72ºF °F 125 gtt15.
Infrared Thermometer The Product 12 13 38 mm @ 300 mm 60 mm @ 100 mm @ 600 mm 900 mm 33 mm @ 300 mm 50 mm @ 600 mm 85 mm @ 900 mm S 1.5" @ 12" 2.4" @ 24" D 62 MAX D:S = 10:1 4" @ 36" S 1.4" @ 12" 2" @ 24" 3.4" @ 36" D 62 MAX + D:S = 12:1 gtt08-11.
62 MAX/62 MAX + Users Manual 14 70 21 gtt03.
Infrared Thermometer The Product 15 75° 225° 75° gtt04.
62 MAX/62 MAX + Users Manual 16 17 + gtt13.