NetDAQ Users Manual
Table Of Contents
- 2640A/2645A NetDAQ Users Manual
- 1. Overview
- 2. Preparing for Operation
- Introduction
- Instrument Preparation
- Unpacking and Inspecting the Instrument
- Positioning and Rack Mounting
- Connecting to a Power Source and Grounding
- Universal Input Module Connections
- Digital I/O Connections
- Alarm/Trigger I/O Connections
- External Trigger Wiring for a Group Instrument
- Controls and Indicators
- Front Panel Operating Procedures
- Power-On Options
- Displaying a Monitor Channel
- Displaying the Digital I/O Status
- Displaying the Totalizer Status
- Reviewing and Setting the Base Channel Number
- Reviewing and Setting the Line Frequency
- Reviewing and Setting the Network Type
- Reviewing and Setting the General Network Socket Port
- Reviewing and Setting the General Network IP Address
- Reviewing and Setting the Subnet Mask and Default Gateway
- Viewing the Instrument Ethernet Address
- Host Computer and Network Preparation
- Testing and Troubleshooting
- 3. Configuring NetDAQ Logger for Windows
- Introduction
- Configuring Network Communications
- Configuring the Current Setup
- Setup Files
- Configuring an Instrument
- Configuring Channels
- Configuring Mx+B Scaling From a File
- Entering an Instrument's Description
- Copying a Channels Configuration
- Default Configuration Settings
- Using Configuration Lockout
- Saving an Instrument's Configuration as a Text File
- Configuring the netdaq.ini File
- 4. Operating NetDAQ Logger for
- 5. Using Trend Link for Fluke
- Introduction
- Getting the Right Look for Your Trend Link Chart
- 6. Maintenance
- Introduction
- Self-Test Diagnostics and Error Codes
- Cleaning
- Fuse Replacement
- Performance Test
- Configuring the Performance Test Setup
- Initializing the Performance Test Setup
- Accuracy Performance Tests
- Volts DC Accuracy Test (2640A)
- Volts DC Accuracy Test (2645A)
- Volts AC Accuracy Test
- Frequency Accuracy Test
- Analog Channel Integrity Test
- Computed Channel Integrity Test
- Thermocouple Temperature Accuracy Test
- Open Thermocouple Response Test
- 2-Wire Resistance Accuracy Test (2640A)
- 2-Wire Resistance Accuracy Test (2645A)
- 4-Wire Resistance Accuracy Test (2640A)
- 4-Wire Resistance Accuracy Test (2645A)
- RTD Temperature Accuracy Test (Resistance) (2640A)
- RTD Temperature Accuracy Test (Resistance) (2645A)
- RTD Temperature Accuracy Test (DIN/IEC 751 RTD)
- Digital Input/Output Tests
- Totalizer Tests
- Master Alarm Output Test
- Trigger Input Test
- Trigger Output Test
- Calibration
- Variations in the Display
- Service
- Replacement Parts
- Appendices
- A. Specifications
- Introduction
- 2640A/2645A Combined Specifications
- 2640A Specifications
- 2640A DC Voltage Measurement Specifications
- 2640A AC Voltage Measurement Specifications
- 2640A 4-Wire Resistance Measurement Specifications
- 2640A 2-Wire Resistance Measurement Specifications
- 2640A RTD's 4-Wire, per ITS-1990 Measurement Specifications
- 2640A RTD's 2-Wire per ITS-1990 Measurement Specifications
- 2640A Thermocouple per ITS-1990 Measurement Specifications
- 2640A Frequency Measurement Specifications
- 2645A Specifications
- 2645A DC Voltage Measurement Specifications
- 2645A AC Voltage Measurement Specifications
- 2645A 4-Wire Resistance Measurement Specifications
- 2645A 2-Wire Resistance Measurement Specifications
- 2645A 4-Wire RTD per ITS-1990 Measurement Specifications
- 2645A Thermocouple per ITS-1990 Measurement Specifications
- 2645A Frequency Measurement Specifications
- B. Noise, Shielding, and Crosstalk Considerations
- C. True-RMS Measurements
- D. RTD Linearization
- E. Computed Channel Equations
- F. Data File Format
- G. Dynamic Data Exchange (DDE)
- H. Ethernet Cabling
- I. Network Considerations
- J. Error Messages & Exception Conditions
- K. Fluke Service Centers
- A. Specifications
- Index
- Instrument Parameter Record (Isolated Network)
- Instrument Parameter Record (General Network)
- General Network Parameter Record
- Host Computer General Network Parameter Record

2640A/2645A NetDAQ
Users Manual
6-16
Thermocouple Temperature Accuracy Test 6-15.
Ensure that the Accuracy Tests (above) have been completed before performing
this test.
1. Connect a Thermocouple Remove the Universal Input Module from the
instrument and connect the supplied type T thermocouple to the channel 1
terminals with the blue lead to the H terminal and red lead to the L terminal.
Reinstall the Universal Input Module.
2. Configure Channel 1 for Thermocouples In NetDAQ Logger, configure
channel 1 for Thermocouples with Range (thermocouple type) T. (See
“Configuring Analog Channel Functions" in Chapter 3 of this manual.)
3. Open Spy Window Select the Spy command from the Utilities menu. Select
analog channel 01. Click OK to open the Spy window.
4. Verify Accuracy Insert the thermocouple and a mercury thermometer in a
room-temperature bath. Allow 20 minutes for thermal stabilization. The value
displayed on the mercury thermometer should equal the value in the Spy
Window + 0.5°C (2640A) or +
0.95°C (2645A) plus any sensor inaccuracies.
5. Close Spy Window.
Open Thermocouple Response Test 6-16.
This test checks the Open Thermocouple response.
1. Connect an 820 Ohm Test Resistor Remove the Universal Input Module
from the instrument and connect an 820 ohm resistor to the channel 1
terminals. Reinstall the Universal Input Module.
2. Configure Channel 1 for Thermocouples In NetDAQ Logger, configure
channel 1 for Thermocouples with Range (thermocouple type) K. (See
“Configuring Analog Channel Functions" in Chapter 3 of this manual.)
3. Open Spy Window Select the Spy command from the Utilities menu. Select
analog channel 1. Click OK to open the Spy window. The value displayed
should approximate the ambient temperature.
4. Connect a 10k Ohm Test Resistor Remove the Universal Input Module
from the instrument and connect a 10k ohm resistor to the channel 1 terminals
to simulate an open thermocouple condition. Reinstall the Universal Input
Module.
5. Verify Open Thermocouple The Spy window indicates an open
thermocouple detect condition by displaying Open TC in place of a
temperature reading.
6. Close Spy Window.