NetDAQ Users Manual
Table Of Contents
- 2640A/2645A NetDAQ Users Manual
- 1. Overview
- 2. Preparing for Operation
- Introduction
- Instrument Preparation
- Unpacking and Inspecting the Instrument
- Positioning and Rack Mounting
- Connecting to a Power Source and Grounding
- Universal Input Module Connections
- Digital I/O Connections
- Alarm/Trigger I/O Connections
- External Trigger Wiring for a Group Instrument
- Controls and Indicators
- Front Panel Operating Procedures
- Power-On Options
- Displaying a Monitor Channel
- Displaying the Digital I/O Status
- Displaying the Totalizer Status
- Reviewing and Setting the Base Channel Number
- Reviewing and Setting the Line Frequency
- Reviewing and Setting the Network Type
- Reviewing and Setting the General Network Socket Port
- Reviewing and Setting the General Network IP Address
- Reviewing and Setting the Subnet Mask and Default Gateway
- Viewing the Instrument Ethernet Address
- Host Computer and Network Preparation
- Testing and Troubleshooting
- 3. Configuring NetDAQ Logger for Windows
- Introduction
- Configuring Network Communications
- Configuring the Current Setup
- Setup Files
- Configuring an Instrument
- Configuring Channels
- Configuring Mx+B Scaling From a File
- Entering an Instrument's Description
- Copying a Channels Configuration
- Default Configuration Settings
- Using Configuration Lockout
- Saving an Instrument's Configuration as a Text File
- Configuring the netdaq.ini File
- 4. Operating NetDAQ Logger for
- 5. Using Trend Link for Fluke
- Introduction
- Getting the Right Look for Your Trend Link Chart
- 6. Maintenance
- Introduction
- Self-Test Diagnostics and Error Codes
- Cleaning
- Fuse Replacement
- Performance Test
- Configuring the Performance Test Setup
- Initializing the Performance Test Setup
- Accuracy Performance Tests
- Volts DC Accuracy Test (2640A)
- Volts DC Accuracy Test (2645A)
- Volts AC Accuracy Test
- Frequency Accuracy Test
- Analog Channel Integrity Test
- Computed Channel Integrity Test
- Thermocouple Temperature Accuracy Test
- Open Thermocouple Response Test
- 2-Wire Resistance Accuracy Test (2640A)
- 2-Wire Resistance Accuracy Test (2645A)
- 4-Wire Resistance Accuracy Test (2640A)
- 4-Wire Resistance Accuracy Test (2645A)
- RTD Temperature Accuracy Test (Resistance) (2640A)
- RTD Temperature Accuracy Test (Resistance) (2645A)
- RTD Temperature Accuracy Test (DIN/IEC 751 RTD)
- Digital Input/Output Tests
- Totalizer Tests
- Master Alarm Output Test
- Trigger Input Test
- Trigger Output Test
- Calibration
- Variations in the Display
- Service
- Replacement Parts
- Appendices
- A. Specifications
- Introduction
- 2640A/2645A Combined Specifications
- 2640A Specifications
- 2640A DC Voltage Measurement Specifications
- 2640A AC Voltage Measurement Specifications
- 2640A 4-Wire Resistance Measurement Specifications
- 2640A 2-Wire Resistance Measurement Specifications
- 2640A RTD's 4-Wire, per ITS-1990 Measurement Specifications
- 2640A RTD's 2-Wire per ITS-1990 Measurement Specifications
- 2640A Thermocouple per ITS-1990 Measurement Specifications
- 2640A Frequency Measurement Specifications
- 2645A Specifications
- 2645A DC Voltage Measurement Specifications
- 2645A AC Voltage Measurement Specifications
- 2645A 4-Wire Resistance Measurement Specifications
- 2645A 2-Wire Resistance Measurement Specifications
- 2645A 4-Wire RTD per ITS-1990 Measurement Specifications
- 2645A Thermocouple per ITS-1990 Measurement Specifications
- 2645A Frequency Measurement Specifications
- B. Noise, Shielding, and Crosstalk Considerations
- C. True-RMS Measurements
- D. RTD Linearization
- E. Computed Channel Equations
- F. Data File Format
- G. Dynamic Data Exchange (DDE)
- H. Ethernet Cabling
- I. Network Considerations
- J. Error Messages & Exception Conditions
- K. Fluke Service Centers
- A. Specifications
- Index
- Instrument Parameter Record (Isolated Network)
- Instrument Parameter Record (General Network)
- General Network Parameter Record
- Host Computer General Network Parameter Record

Specifications
2645A Specifications
A
A-21
2645A 4-Wire Resistance Measurement Specifications A-24.
Tables A-30 to A-32 provide 2645A specifications for the 4-wire resistance
measurement function. The 4-wire measurements use 2 input channels a decade
apart, e.g., channels 4 and 14.
Table A-30. 2645A 4-Wire Resistance Temperature Coefficient
Specification Characteristic
Temperature Coefficient Add 1/10th the 90 day specification per ºC
above 28ºC or below 18ºC.
Accuracy at -20ºC Multiply the -10°C to + 60°C accuracy
specification by 2. After 1 hour warm-up. For
accuracy between -10°C and -20°C, interpolate
linearly.
Table A-31. 2645A 4-Wire Resistance Range and Resolution Specifications
Range
Resolution
Current
Applied
Full Scale
Voltage
Maximum
Voltage
Slow Fast
Applied
300Ω 10 mΩ 30 mΩ 1 mA 300 mV 3.5V
3 kΩ 100 mΩ 300 mΩ 100 µA 300 mV 3.5V
30 kΩ 1Ω 3Ω 10 µA 300 mV 3.5V
300 kΩ 10Ω 30Ω 10 µA 3.0V 3.5V
3 MΩ 100Ω 300Ω 1 µA 3.0V 3.5V
Table A-32. 2645A 4-Wire Resistance Accuracy Specifications
Accuracy, 3σ ± (% input + V)
18ºC to 28ºC-10ºC to 60ºC
Range
90 Day 1 Year 1 Year
Slow Fast Slow Fast Slow Fast
300Ω .02%+60 mΩ .02%+.1Ω .02%+.1Ω .02%+.2Ω .084%+.25Ω .084%+.42Ω
3 kΩ .02%+.6Ω .02%+2Ω .02%+1Ω .02%+3Ω .084%+2.5Ω .084%+8.4Ω
30 kΩ .02%+6Ω .2%+200Ω .02%+10Ω .2%+300Ω .084%+25Ω .84%+840Ω
300 kΩ .5%+80Ω 1%+2 kΩ .5%+150Ω 1%+3 kΩ 2.1%+336Ω 4.2%+8.4 kΩ
3 MΩ 1.3%+1 kΩ 2%+120 kΩ 1.3%+2 kΩ 2%+200 kΩ 5.46%+4.2 kΩ 8.4%+200 kΩ