Datasheet
Table Of Contents

i400
Calibration Information
6
Current Accuracy Tests
1. Connect the equipment as shown in Figure 2.
2. Center the UUT jaws around the single-turn #14 gauge wire.
3. Set the 5520A output as called out in Table 3. Check that the Multimeter readings are within the
specified limits in step 1 of Table 3.
4. Place the 5520A in Standby mode.
5. Connect the equipment as shown in Figure 3.
6. Center the jaws of the UUT around the 5500A coil.
7. Apply the 5520A output as called out in Table 2, steps 2 and 3. Verify that the Multimeter readings are
within the specified limits.
Table 3. Accuracy Tests
*Multimeter Reading limits
Step Nominal Value 5520A Output
5520A LCOMP
Setting
UUT Nominal
Output
Lower Upper
1. 1 A, 45 Hz 1 A, 45 Hz OFF 1 mA, 45 Hz 0.920 mA 1.080 mA
2. 100 A, 400 Hz 2.0 A, 400 Hz ON 100 mA, 400 Hz 97.94 mA 102.06 mA
3. 400 A, 60 Hz 8.0 A, 60 Hz ON 400 mA, 60 Hz 391.94 mA 408.06 mA
*To obtain the best TUR, the Fluke 189 is pre-calibrated at each step using the 5520A with LCOMP OFF. The Fluke 189
error is then algebraically added to the UUT reading. The Fluke 189 must be in a mA B function.
MIN MAXHOLD
REL
%
ms
Hz
RANGE
dB
dB
ac+dc
ac+dc
ac+dc
ac+dc
F
nS
mA
mA
A
mV
V
mV
V
OFF
C
A
A
A
A
mA
COM
V
TEMPERATURE
A
MIN MAXHOLD
REL
%
ms
Hz
RANGE
dB
dB
ac+dc
ac+dc
ac+dc
ac+dc
F
nS
mA
mA
A
mV
V
mV
V
OFF
C
A
A
A
A
mA
COM
V
TEMPERATURE
A
AutoHOLD
LOGGING
SAVECANCEL
FAST MN MX
SETUP
YES
NO
CLEAR MEM
VIEW
LOGGING MULTIMETER
189
bbs04f.eps
Figure 2. Connections for Current-Accuracy Test (Step 1)