User Manual

®
i200/i200s
AC Current Clamp
Instruction Sheet
Introducing the i200/i200s
The i200 is a single range 200A clamp-on AC Current Clamp with
current output via safety shrouded banana plugs.
The i200s is a dual range 20A and 200A clamp-on AC Current
Clamp with voltage output via a safety insulated BNC connector.
A dual banana to BNC adapter is supplied to allow the i200s to be
connected to multimeters with banana input.
Unpacking
The following items should be included in your Current Clamp box:
Current Clamp
Dual Banana to BNC Adapter model PM9081 (only with i200s)
Instruction Sheet (this paper)
Check the contents of the shipping box for completeness. If
something in the box has been damaged or missing, contact your
distributor or the nearest FLUKE sales or service office
immediately.
Using the Current Clamp Safely
Warning
To prevent electric shock or fire and personal injury,
carefully read all safety information before attempting
to operate the Current Clamp and follow these
procedures:
Do not use the clamp on circuits rated higher than 600V in
Installation Category III (See "Specifications", subsection
“SAFETY”.) Use caution when clamping around uninsulated
conductors or bus bars.
Do not use a clamp that is cracked, damaged, or has a
defective cable. Such clamps should be made inoperative by
taping the clamp shut to prevent operation.
Check the magnetic mating surfaces of the clamp jaws; these
should be free of dust, dirt, rust, and other foreign matter.
Keep your fingers off the clamp jaws.
Keep your fingers behind the safety barrier as shown in
figure1.
Figure 1. Safely holding the Current Clamp
In this Users Manual, a Warning identifies conditions and actions
that pose hazard(s) to the user. A Caution identifies conditions and
actions that may damage the current clamp. International electrical
symbols used are explained below.
See explanation in
manual
Conformité
Européenne
Double Insulation
(Protection class)
Earth
Specifications
SAFETY
Input jaws
and
output floating voltage to
ground
Meets requirement of:
EN/IEC61010-1
EN/IEC61010-2-032
UL3111,
for 600V CAT III,
pollution degree 2.
CE Conformity Meets requirement of:
EN 50081-1, EN 50082-2
ELECTRICAL SPECIFICATIONS
All Electrical Specifications are valid at the following reference
conditions:
Ambient temperature
23±3°C (73±3°F)
Relative Humidity
20 to 75%
Frequency
48 to 65 Hz
Continuous external field
< 40 A/m
Load impedance
i200: 0.2
…15
i200s: >1 M
// 100 pF
The current may not contain any DC component
No influence from adjacent currents
The conductor must be centered within the jaw aperture
20A Range (i200s only)
Measuring range 0.1 to 24A
Maximum current 24A
Crest factor *) < 3
Maximum non-destructive
current
200A (Frequency 1 kHz and
crest factor < 3)
Output signal 100 mV/A
Output impedance
20 @ 1 kHz
Basic accuracy
48 Hz to 65 Hz
2% + 0.5A
Additional error:
40 Hz to 48 Hz and
65 Hz to 1 kHz
+ < 10%
1 kHz to 10 kHz + < 15%
Phase shift Unspecified
200A Range i200 i200s
Measuring range 0.5 to 240A 0.5 to 240A
Maximum current 240A 240A
Crest factor < 3 < 3
Maximum non-destructive
current
@ Frequency 1 kHz and
crest factor < 3
Continuous 200A
10 min ON
/30 min OFF
240A
Output signal 1 mA/A 10 mV/A
Output impedance -
10 @ 1 kHz
Basic accuracy
48 Hz to 65 Hz
0.5A to 10A
3% + 0.5A 3.5% + 0.5A
10A to 40A
2.5% + 0.5A 3% + 0.5A
40A to 100A
2% + 0.5A 2.5% + 0.5A
100A to 240A
1% + 0.5A 1.5% + 0.5A
Additional error:
40 Hz to 48 Hz and
65 Hz to 1 kHz
+ < 3%
+ < 3%
1 kHz to 10 kHz + < 12% + < 12%
Phase shift
0.5A to 10A Unspecified Unspecified
10A to 40A
5 ° 6 °
40A to 100A
3 ° 4 °
100A to 240A
2.5 ° 3 °
All ranges i200 i200s
Load on output
0.2...15 >1 M // < 100 pF
Load Influence Current: < 1%
Phase: < 1°
-
-
Bandwidth -1.5 dB
-3dB
40 Hz to 10 kHz
40 kHz
40 Hz to 10 kHz
40 kHz
Additional errors
With temperature
0.15 % / 10 K
With position of
conductor in the
clamp aperture
0.5 % @ 50 Hz
With adjacent
conductors
15 mA / A @ 50 Hz
*) This is the maximum permissible ratio between the peak value
of a superimposed transient and the ac rms value.
GENERAL
Clamp Dimensions 135 x 50 x 30 mm
(5.3 x 2 x 1.2 in)
Protection index IP40
Jaw Opening 21 mm (0.82 in)
Height of open Jaws 69 mm (2.7 in)
Maximum conductor size
20 mm (0.8 in) or
busbar 20 x 5 mm (0.8 x 0.2 in)
Weight 180 g (6.4 oz)
Cable length i200
i200s
1.5 m (5.9 in)
2m (7.9 in)
Temperature
Operating
Non-operating
-10 to +55°C (+14 to +131°F)
-40 to + 70°C (-40 to +158°F)
Relative Humidity
Operating
85%, up to +30°C (+86°F)
75%, up to +55°C (+131°F)
Altitude
Operating
Non-operating
to 2000 m (6500 ft)
to 12000 m (40000 ft)
EMC EN 50081-1
EN 50082-2 (3V/m, 2.74V/yd)
Instrument Compatibility
The i200s is compatible with any Fluke ScopeMeter test tool,
Power Harmonics Analyzer, Oscilloscope, Multimeter, or other
voltage measurement device that has the following features:
BNC input connector. The Dual Banana to BNC Adapter
included in the package, can be used to connect to standard
inputs on multimeters. For the 120 series ScopeMeters, use
the BB120 Shielded Banana to BNC Adapter.
Input accuracy of 2% or better to take full advantage of the
accuracy of the Current Clamp.
Input impedance of greater than or equal to 1 M, and for full
bandwidth and accuracy, a maximum input capacity of 100 pF.
A passband of more than four times the frequency of the
waveform to be measured.
The i200 is compatible with any Fluke Multimeter or any other
current measurement device that has the following features:
Banana inputs.
Input accuracy of 2% or better to take full advantage of the
accuracy of the Current Clamp.
Input impedance of 0.2 …15
A passband of more than four times the frequency of the
waveform to be measured.
PN 4822 872 00934