User Manual
8808A Users Manual Supplement
8/08 1
Change #1, 44931, 47892
On page 1-8, under Relative Humidity (non-condensing), replace Operating with the
following:
Operating…………………………………<90 % (0 °C to 28 °C)
<75 % (28 °C to 40 °C)
<45 % (40 °C to 50 °C)
Change #2, 46479
On page 1-6, Table 1-2, change the description for >,
From: Standby power ON/OFF
To: Display ON/OFF and Meter reset
On page 1-11, under DC Current, Shunt Resistance, replace the Burden Voltage with the
following:
Burden voltage <5 mV for 200 µA and 2 mA range.
Under Input Characteristics, change the Burden Voltage for both 200 µA and 2 mA
ranges,
From: <1 mV
To: <5 mV
On page 2-5, under the Warning, replace the second sentence in step 4 with the following:
If the fuse is good, the Meter displays a reading of <10 Ω.
On page 3-9, under Selecting a Measurement Rate, change the order of the last two
sentences in the last paragraph as follows:
Pressing 8 does not affect the frequency update rate. The measurement rate is always fast for
Diode and Continuity test.
On page 3-11 under Frequency Ranging, in the third paragraph, delete the second sentence:
If the frequency is lower than 20 Hz, =9is displayed.
On page 3-20, change the following and delete Level 4:
From:
• Level 1 (0.01 % of reading)
• Level 2 (0.1 % of reading)
• Level 3 (1 % of reading)
• Level 4 (10 % of reading)
To:
• Level 1 (5 % of range)
• Level 2 (7 % of range)
• Level 3 (8 % of range)
On page 3-25, Table 3-9, replace the 4
th
row with the following:
Touch Hold Sensitivity level 1 (5 % of reading)