User Manual
  Digital Multimeter 
  Detailed Specifications 
 43 
Temperature 
Range Resolution 
Accuracy 
[1,2]
-200 °C to +1090 °C 
-328 °F to +1994 °F 
0.1 °C 
0.1 °F 
±(1.0 % + 10) 
±(1.0 % + 18) 
[1]  Does not include error of the thermocouple probe. 
[2]  Accuracy specification assumes ambient temperature stable to ± 1 °C. For ambient temperature changes of ± 5 °C, rated accuracy 
applies after 2 hours. 
AC Current 
Function Range  Resolution Burden Voltage  Accuracy 
[1]
μA ac 
600.0 μA 0.1 μA 100 μV/μA 
±(1.5 % + 2)  ±(1.0 % + 2) 
6000 μA 1 μA 100 μV/μA 
mA ac 
60.00 mA  0.01 mA  1.8 mV/mA 
400.0 mA
 [2]
0.1 mA  1.8 mV/mA 
A ac 
6.000 A  0.001 A  0.03 V/A 
10.00 A
 [3,4]
0.01 A  0.03 V/A 
[1]  AC conversions are ac coupled, true rms responding, and valid from 3 % to 100 % of range, except 400 mA range. (5 % to 100 % of 
range) and 10 A range (15 % to 100 % or range). 
[2]  400 mA continuous. 600 mA for 18 hr maximum. 
[3]  W 10 A continuous up to 35 °C. <20 minutes on, 5 minutes off at 35 °C to 55 °C. >10 A to 20 A for 30 seconds maximum, 5 minutes 
off. 
[4]  >10 A accuracy unspecified. 