34
Touch top half
of input contacts
Good 11 A fuse: 00.0 Ω to
00.5 Ω
Replace fuse: OL
Good 0.44 A fuse: 0.995 kΩ to
1.005 kΩ
Replace fuse: OL
gaq105.emf
Figure 10. Current Fuse Test
How to Replace the Batteries
Replace the batteries with three AA batteries
(NEDA 15A IEC LR6).
XWWarning
To prevent possible electrical shock, fire, or
personal injury:
Batteries contain hazardous chemicals
that can cause burns or explode. If
exposure to chemicals occurs, clean
with water and get medical aid.
Repair the Product before use if the
battery leaks. Battery leakage may create
a shock hazard or damage the Product.
Do not put battery cells and battery
packs near heat or fire. Do not put in
sunlight.
Replace the battery as follows, refer to Figure 11:
1. Turn the rotary switch to OFF and remove the test
leads from the terminals.
2. Remove the six Phillips-head screws from the case
bottom and remove the battery door ().
Note
While lifting the battery door, ensure the rubber
gasket stays attached to the battery
compartment barrier.