Datasheet

time. Using on-board TrendCapture users can
graphically review logged readings without
needing a PC. Store up to 15,000 recorded
events.
Save multiple logging sessions before pc down-
load is necessary
Two terminal 50 ohm range with 1 milliohm
resolution, 10 mA source current. Useful for
measuring and comparing differences in motor
winding resistance or contact resistance.
Low Pass filter for accurate voltage and fre-
quency measurements at the same time on
adjustable speed motor drives and other elec-
trically noisy equipment
LoZ Volts. Low impedance voltage function for
eliminating ghost voltages. Also recommended
when testing for absence or presence of live
power.
button. On board help screens for measure-
ment functions. Unsure about a function—go to
that function and press the “i” button.
Real time clock for automatic time stamping of
saved readings
Relative mode to remove test lead resistance
from low ohms or capacitance measurements
True-rms ac voltage and current for accurate
measurements on non linear signals
Measure up to 10 A (20 A for 30 seconds)
100 mF capacitance range
Peak capture to record transients as fast as
250 µs
Premium test leads and alligator clips included
Amp jack plugs included
Optional magnetic hanger for easy setup and
viewing while freeing your hands for other
tasks
Optional FlukeView forms enables you to docu-
ment, store and analyze individual readings or
a series of measurements, then convert them
into professional-looking documents
Limited lifetime warranty
2 Fluke Corporation 289 True-rms Industrial Logging Multimeter with TrendCapture
Specifications
Function Range and Resolution Basic Accuracy
DC volts
AC volts
50.000 mV, 500.00 mV, 5.0000 V, 50.000 V,
500.00 V, 1000.0 V
0.025 %
0.4 % (true-rms)
DC current
AC current
500.00 µA, 5000.0 µA, 50.000 mA, 400.00 mA,
5.0000 A, 10.000 A
0.05 %
0.6 % (true-rms)
Temperature
(excluding probe)
-200.0 °C to 1350.0 °C (-328.0 °F to 2462.0 °F) 1.0 %
Resistance 500.00 Ω, 5.0000 kΩ, 50.000 kΩ, 500.00 kΩ,5.0000 MΩ, 50.00 MΩ, 500.0 MΩ 0.05 %
Capacitance 1.000 nF,10.00 nF 100.0 nF, 1.000 µF, 10.00 µF, 100.0 µF, 1000 µF, 10.00 mF,
100 mF
1.0 %
Frequency 99.999 Hz, 999.99 Hz, 9.9999 kHz, 99.999 kHz, 999.99 kHz 0.005 %
Connectivity Optional infrared connector via Fluke ir3000 FC
Additional functions/features Fluke 289
Multiple on screen displays Yes
True-rms ac bandwidth 100 kHz
dBV/dBm Yes
DC mV resolution 1 µV
Megohm range up to 500 MΩ
Conductance 50.00 nS
Continuity beeper Yes
Battery/fuse access Battery/fuse
Elapse time clock Yes
Time of day clock Yes
Min-max-avg Yes
Duty cycle
Yes
Pulse width
Yes
Isolated optical interface Yes
Auto/touch hold Yes
Reading memory Yes
Log to PC Yes
Interval/event logging Yes
Logging memory up to 15,000 readings