User Manual
  True-rms Remote Display Digital Multimeter 
  Detailed Specifications 
41 
AC Current 
Function Range  Resolution 
Accuracy 
(45 – 500 Hz) 
6.000 A  0.001 A 
A ac
 [1,2,3]
10.00 A  0.01 A 
±(1.5 % + 3) 
[1]  All ranges are specified from 5 % of range to 100 % of range. 
[2]  Crest factor of ≤3 at 4000 counts, decreasing linearly to 1.5 at full scale. 
[3]  AC current >10 A is unspecified. 20 A continuous overload for 30 seconds maximum. 
DC Current 
Function Range  Resolution  Accuracy 
6.000 A  0.001 A 
A dc
 [1]
10.00 A  0.01 A 
±(1.0 % + 3) 
[1]  DC current >10 A is unspecified. 20 A continuous overload for 30 seconds maximum. 