User Manual
True-rms Multimeters
Electrical Specifications
23
Function Overload Protection
[1]
Input Impedance
(Nominal)
Common Mode Rejection Ratio
(1 kΩ Unbalanced)
Normal Mode Rejection
Volts ac 1000 V rms >10 MΩ < 100 pF >60 dB @ dc, 50 Hz or 60 Hz
Volts dc 1000 V rms >10 MΩ < 100 pF >120 dB @ dc, 50 Hz or 60 Hz >60 dB @ 50 Hz or 60 Hz
mV/ 1000 V rms
[2]
>10 MΩ < 100 pF >120 dB @ dc, 50 Hz or 60 Hz >60 dB @ 50 Hz or 60 Hz
Open Circuit Test
Voltage
Full Scale Voltage To:
Short Circuit Current
600 kΩ 50 MΩ
Ohms/Capacitance 1000 V rms
[2]
<8.0 V dc <660 mV dc <4.6 V dc <1.1 mA
Continuity/Diode test 1000 V rms
[2]
<8.0 V dc 2.4 V dc <1.1 mA
[1] 10
7
V-Hz maximum.
[2] For circuits <0.3 A short circuit. 660 V for high energy circuits.
Function Overload Protection Overload
mA Fused, 44/100 A, 1000 V FAST Fuse 600 mA overload for 2 minutes maximum, 10 minutes rest minimum
A Fused, 11 A, 1000 V FAST Fuse 20 A overload for 30 seconds maximum, 10 minutes rest minimum