User Manual
1587 FC/1587/1577
Calibration Manual
34
Table 7. Calibration Adjustment Steps for 1587 FC/1587 Models with SN >32980000
Switch Position
(Function)
Input
Terminal
Calibration Adjustment Step Input Value
AC Volts V/COM 600.0 mV, 60 Hz
600.0 mV, 5 kHz
6.000 V, 60 Hz
6.000 V, 5 kHz
60.00 V, 60 Hz
60.0 V, 5 kHz
600.0 V, 60 Hz
600.0 V, 2 kHz
DC Volts 6.000 V, 0 Hz
60.00 V, 0 Hz
600.0 V, 0 Hz
DC Millivolts 600.0 mV, 0 Hz
60.00 mV, 0 Hz
Ohms 600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
0.000 Ω
50.0 MΩ
Continuity 600.0 Ω
Diode Test 6.000 V, 0 Hz
milliamps mA/COM 60.00 mA, 60 Hz
400.0 mA, 60 Hz
60.00 mA, 0 Hz
400.0 mA, 0 Hz
Insulation mA/COM 1.0 mA, 0 Hz
XWWarning
To avoid electrical shock hazard, remove all test leads from
the UUT before performing the following step.
Insulation None
[1]
None (press BLUE to
start internal cal)
Notes:
[1] Internal HV output calibration. Ensure no lead is inserted in Insulation (+) jack.