User Manual

1587 FC/1587/1577
Calibration Manual
32
Meter Keys Used in the Calibration Steps
The Meter keys behave as follows when performing the Calibration Adjustment
Procedure. This may be of help determining why a calibration step is not
accepted and for determining the input value without referring to Table 6.
H Press and hold to show the measured value. The measurement value
is not calibrated so it may not match the input value. This is normal.
Press and hold to display the required input amplitude.
Press and hold to display the frequency of the required input.
G Store the calibration value and advance to the next step. This key is
also used to exit the calibration mode after the calibration-adjustment
sequence is complete.
Calibration Adjustment
Use the following steps to make calibration adjustments to the UUT. Complete
the adjustment procedure before turning the UUT off; otherwise, the new
calibration constants will not be saved.
1. While holding down , turn the rotary switch from OFF to F. The UUT
displays Z .
2. Press once to see the calibration counter, for example, 0001.
3. Press again to start the password entry. The UUT displays ????.
4. Press four keys to enter the password.
5. Press to go to the first calibration step. The UUT displays 0001 if the
password is correct. If the password is not correct, the UUT emits a double
beep, displays ???? and the password must be entered again. Repeat step 4.
6. Use Table 6, Table 7, or Table 8 according to the serial number and model of
your Meter. Apply the input value listed for each calibration adjustment step.
For each step, position the rotary switch and apply the input to the terminals
as indicated in the table.
7. After each input value is applied, press G to accept the value and proceed to
the next step (0002 and so forth).
Note
After pressing G, wait until the step number advances before
changing the calibrator source or turning the rotary switch. If the
rotary switch is not in the correct position, or if the measured value is
not within the anticipated range of the input value, the UUT emits a
double beep and will not continue to the next step. Some adjustment
steps take longer to execute than others (10 to 15 seconds). For
these steps, the UUT will beep when the step is complete. Not all
steps have this feature.
8. After the final step, the display shows  to indicate that the calibration
adjustment is complete. Press G to go to meter mode.