User Manual
  Industrial ScopeMeter® 
  Maintenance 
63 
AB
BB120
1
hxv53.eps 
Figure 14. 10:1 Scope Probes 
6. Use 
YZ to highlight Probe Adjust. 
7. Push  to open the PROBE ADJUST menu. 
A square wave appears on the screen. 
8.  Adjust the trimmer screw in the probe housing to 
give an optimum square wave. 
9. Push 4 to exit the menu. 
Calibration Information 
The Test Tool specifications are based on a 1 year 
calibration cycle. Recalibration must be done by qualified 
personnel. Contact your local Fluke representative for 
more information about recalibration. 
To find the firmware version and calibration date of your 
Test Tool: 
1. Push  to open the Menu. 
2. Use 
YZ to highlight USER OPTIONS. 
3. Push  to open the USER OPTIONS menu. 
4. Use 
YZ to highlight Information. 