User Manual
  Industrial ScopeMeter® 
  Test Tool Setup 
11 
Measurement Connections 
The Test Tool has two 4 mm safety-shielded banana jack 
inputs (red input A and blue input B) and a black safety 
4 mm banana jack input (COM). See Figure 4. 
hxv05.eps 
Figure 4. Input Connections for Measurement 
Input A 
Use the input A (red) for all single input measurements. 
Input B 
For measurements on two different signals, use B input 
(blue) together with input A (red). 
COM 
Use the COM input (black) as the single ground for low 
frequency, continuity, Ohm (Ω), diode, and capacitance 
measurements. 
XW Warning 
To prevent electrical shock or fire, use only 
one COM ˜ (common) connection, or ensure 
that all connections to COM ˜ are at the 
same potential. 
Measurement Probe Setup 
The Test Tool supports measurement probes, for 
example: 
•  10:1 voltage probe 
• 1 mV/°C temperature probe 
•  10 mV/A current clamp 
To set up a specific probe type: 
1. Push  to open the Scope and Meter menu. 
2. Push 3 to open the INPUT A menu. 
3. Use Y/Z to highlight SELECT. 
4. Push  to open the PROBE select menu. 
5. Use 
Y/Z to highlight the probe type. 
6. Push  to accept the probe type and close the 
menu. 