Datasheet
7 Fluke Corporation Fluke 120B Series Industrial ScopeMeter
®
Hand-Held Oscilloscopes
Temperature (TEMP) with optional temperature probe
Range 200 °C/div (200 °F/div)
Scale factor 1 mV/°C and 1 mV/°F
Accuracy as V dc (add temp. probe uncertainty)
Decibel (dB)
0 dBV 1 V
0 dBm (600 Ω /50 Ω) 1 mW referenced to 600 Ω or 50 Ω
dB on V dc, V ac, or Vac+dc
Full scale reading 1000 counts
Crest factor (CREST)
Range 1 to 10
Full scale reading 90 Counts
Phase
Modes A to B, B to A
Range 0 to 359 degrees
Resolution 1 degree
Power (125B only)
Configurations 1 phase / 3 phase 3 conductor balanced loads (3 phase: fundamental component only, AUTOSET
mode only)
Power factor (PF) Ratio between watts and VA range - 0.00 to 1.00
Watt RMS reading of multiplying corresponding samples of input A (volts) and input B (amperes)
Full scale reading 999 counts
VA Vrms x Arms
Full scale reading 999 counts
VA reactive (var) √((VA)
2
-W
2
)
Full scale reading 999 counts
Vpwm
Purpose to measure on pulse width modulated signals, like motor drive inverter outputs
Principle readings show the effective voltage based on the average value of samples over a whole number of
periods of the fundamental frequency
Accuracy as Vrms for sinewave signals
Input A to common
Ohm (Ω)
Ranges 123B and 124B 500 Ω , 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 30 MΩ
125B 50 Ω, 500 Ω , 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 30 MΩ
Accuracy ± (0.6 % + 5 counts) 50 Ω ±(2 % + 20 counts)
Full scale reading 50 Ω to 5 MΩ - 5000 counts, 30 MΩ - 3000 counts
Measurement current 0.5 mA to 50 nA, decreases with increasing ranges
Open circuit voltage <4 V
Continuity (Cont)
Beep <(30 Ω ± 5 Ω) in 50 Ω range
Measurement current 0.5 mA
Detection of shorts of ≥1 ms
Diode
Measurement voltage @0.5 mA >2.8 V
@open circuit <4 V
Measurement current 0.5 mA
Polarity + on input A, - on COM