•  Do not disassemble or crush battery 
cells and battery packs. 
•  Do not keep cells or batteries in a 
container where the terminals can be 
shorted. 
•  Do not put battery cells and battery 
packs near heat or fire. Do not put in 
sunlight. 
To avoid loss of data, do one of the following before you 
remove the battery pack: 
•  Store the data on a computer or a USB device. 
•  Connect the power adapter. 
To replace the battery pack: 
1.  Turn off the Test Tool. 
2.  Remove all probes and test leads 
3.  Unlock the battery cover. 
4.  Lift the battery cover and remove it from the Test 
Tool. 
5.  Lift one side of the battery pack and remove it from 
the Test Tool. 
6.  Install a good battery pack. 
7.  Place the battery cover into position and lock. 
10:1 Scope Probes 
The 10:1 voltage probe (VP41) that is supplied with the 
Test Tool (varies by model) is always adjusted correctly 
and needs no further adjustment. You need to adjust 
other 10:1 scope probes for optimal response. 
XW Warning 
To prevent possible electrical shock, fire, or 
personal injury, use the BB120 Banana-to-
BNC adapter (delivered with the Test Tool) to 
connect a 10:1 scope probe to the input of 
the Test Tool. 
To adjust probes: 
1.  Connect the 10:1 scope probe from the blue input B 
jack to the red input A jack. 
2.  Use the red 4-mm banana adapter (delivered with 
the probe) and the banana-to-BNC adapter (BB120). 
See Figure 14. 
3. Push  to open the Menu. 
4. Use 
YZ to highlight USER OPTIONS. 
5. Push  to open the USER OPTIONS menu. 