  Industrial ScopeMeter® 
  Fieldbus Mode 
47 
Figure 12 shows the bus health indicator boundaries. The 
high level voltage of a bus must be between +3.0 V (MIN) 
and +15.0 V (MAX). Depending on the measurement 
result, the displayed indicator will be: 
f 
Result is between 4.2 and 13.8V. (10 % of 
12 V = 1.2 V) 
d 
Result is between 3 V and 4.2 V, or between 
13.8 V and 15 V. 
e 
Result is <3 V or >15 V. 
hxv34.eps 
Figure 12. Bus Health Indicator Boundaries 
How to View the Bus Waveform Screen 
To view the waveform eye pattern of the bus voltage: 
1. Push 3. The screen shows the eye pattern. The 
screen shows the waveforms of one bit time 
triggered on a positive, as well as, on a negative 
edge in persistence mode. 
2. Push 1 to clear the persisted waveforms and 
restart to show the waveform. 
hxv35.eps 
3. Push  to freeze the screen. Push  again to 
clear the persistence waveform and restart the 
waveform eye pattern. 