Zooming Harmonics 
If the harmonics bars screen is shown, you can zoom 
vertically for a more detailed view. Use 
YZ to zoom in 
or zoom out. 
The scale at the left side changes when zooming in or 
out. 
Fieldbus Mode 
Fieldbuses are bi-directional, digital, serial control 
networks used in process control and industrial 
automation. 
The Test Tool can indicate the status of the following 
aspects of the OSI model Physical Layer: 
•  Voltage levels (bias, high level, low level) 
•  Bit width – baud rate 
•  Rise and fall time 
• Distortion 
The Test Tool can show the bus signal waveform in the 
Eye-pattern mode, see page 47. 
The Test Tool operates in full automatic (ranging and 
triggering) mode. Test limits are preset, but can be 
changed, see page 47. 
For supported bus types and protocols see Table 9. 
For extended information on Fieldbuses and fieldbus 
measurement, see Appendix A of this manual. 
Note 
To check a suspected cable you can make 
resistance measurements and capacitance 
measurements using the Scope/Meter mode. 
To do fieldbus measurements: 
1. Push  to open the MENU. 
2. Use YZ to highlight BUSHEALTH. 
3. Push  to open the BUS HEALTH menu. 
4. Use 
YZ to highlight the bus type. 
Select User1 or User2 to create a customized set of 
limits to test nonstandard bus systems. See page 48 
for information about how to set the test limits. 
Default settings are RS232 for User1 and Foundation 
Fieldbus H1 for User2. 
5. Push  to make the change. 
For bus types with additional options, a secondary 
menu is available. Use 
YZ to highlight the option 
and  to make the change. 
An example of the screen is shown in Table 10. 
6.  Connect the inputs as shown in Figure 8, setup 4. 