User Manual
  Industrial ScopeMeter® 
  Scope and Meter Mode 
29 
AC Coupling 
Use AC-coupling to observe a small ac signal that rides 
on a dc signal. 
To select ac-coupling on input A: 
1. Push  to open the Scope and Meter button bar. 
2. Push 3 to open the INPUT SETTINGS menu. 
3. Use YZ to highlight the Coupling as AC. 
4. Push  to make the change. 
5. Push 4 to exit the menu. 
Waveform Trigger 
The trigger tells the Test Tool when to start the waveform 
trace. You can: 
•  select which input signal to use 
•  select which edge this should occur 
•  define the condition for a new update of the 
waveform 
The bottom line of the waveform area identifies the trigger 
parameters in use. Trigger icons on the screen indicate 
the trigger level and slope. 
Note 
The trigger settings are fixed for 3-phase power 
measurements. 
Set Trigger Level and Slope 
For quick operation, use  to automatically trigger on 
nearly all signals.  toggles between the Auto and 
Manual mode. The current operating mode is shown in 
the Information Area. 
To optimize trigger level and slope manually: 
1. Repeatedly push  until all menus and 
secondary button bars close. 
2. Push 2 to enable and use WXYZ to set the 
Trigger level and slope adjustment. 