Hart Scientific 9112A Calibration Furnace User’s Guide Rev.
Fluke Corporation, Hart Scientific Division 799 E. Utah Valley Drive • American Fork, UT 84003-9775 • USA Phone: +1.801.763.1600 • Telefax: +1.801.763.1010 E-mail: support@hartscientific.com www.hartscientific.com Subject to change without notice. • Copyright © 2005 • Printed in USA Rev.
Table of Contents 1 Before You Start . . . . . . . . . . . . . . . . . . . . . . . . . . 1 1.1 1.2 Symbols Used . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . 2 1.2.1 1.2.2 1.3 WARNINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Authorized Service Centers. . . . . . . . . . . . . . . . . . . . . .
6.2 Operating the Furnace . . . . . . . . . . . . . . . . . . . . . . . . 23 7 Digital Communication Interface . . . . . . . . . . . . . . . . 25 8 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 9 Appendix A - Material Safety Data Sheets (MSDS) . . . . . .
Figures Figure 1 Figure 2 Figure 3 Figure 4 Front View . . . . . . Sectional Side View . Back View . . . . . . RS-232 Cable Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1 Before You Start Symbols Used 1 1.1 Before You Start Symbols Used Table 1 lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this manual.
9112A Calibration Furnace User’s Guide Symbol Description Canadian Standards Association OVERVOLTAGE (Installation) CATEGORY II, Pollution Degree 2 per IEC1010-1 refers to the level of Impulse Withstand Voltage protection provided. Equipment of OVERVOLTAGE CATEGORY II is energy-consuming equipment to be supplied from the fixed installation. Examples include household, office, and laboratory appliances.
1 Before You Start Safety Information Follow all safety guidelines listed in the user’s manual. Calibration Equipment should only be used by Trained Personnel. If this equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired or safety hazards may arise. Inspect the instrument for damage before each use. DO NOT use the instrument if it appears damaged or operates abnormally.
112A Calibration Furnace User’s Guide ELECTRICAL HAZARD These guidelines must be followed to ensure that the safety mechanisms in this instrument will operate properly. This instrument must be plugged into a 230 VAC (± 10%) 50/60 Hz only electric outlet. The power cord of the instrument is equipped with a three-pronged grounding plug for your protection against electrical shock hazards. It must be plugged directly into a properly grounded three-prong receptacle.
1 Before You Start Authorized Service Centers The unit is a precision instrument. Although it has been designed for optimum durability and trouble free operation, it must be handled with care. Most probes have handle temperature limits. Be sure that the probe handle temperature limit is not exceeded in the air above the instrument. The instrument and any thermometer probes used with it are sensitive instruments that can be easily damaged. Always handle these devices with care.
9112A Calibration Furnace User’s Guide Hart Scientific, Inc. 799 E. Utah Valley Drive American Fork, UT 84003-9775 USA Phone: +1.801.763.1600 Telefax: +1.801.763.1010 E-mail: support@hartscientific.com Fluke Nederland B.V. Customer Support Services Science Park Eindhoven 5108 5692 EC Son NETHERLANDS Phone: +31-402-675300 Telefax: +31-402-675321 E-mail: ServiceDesk@fluke.
1 Before You Start Authorized Service Centers Phone: +65 6799-5588 Telefax: +65 6799-5588 E-mail: antng@singa.fluke.
2 Introduction 2 Introduction The 9112A Calibration Furnace was designed specifically for calibrating PRTs, fiber optic sensors and thermocouples at higher temperature ranges up to 1100°C. The furnace utilizes an equilibration block capable of making comparison measurements on multiple probes. The standard equilibration block is sized for ¼ inch probes, however, custom options are possible. Temperature stability is better than ± 0.
3 Specifications and Environmental Conditions Specifications 3 3.1 3.2 Specifications and Environmental Conditions Specifications Operating Range 300°C to 1100°C Stability 300°C 500°C 700°C 1000°C 1100°C ±0.05°C ±0.05°C ±0.1°C ±0.1°C ±0.1°C Uniformity 300°C 500°C 700°C 1000°C 1100°C ±0.05°C ±0.08°C ±0.2°C ±0.25°C ±0.
9112A Calibration Furnace User’s Guide for a period as stated in our current product catalog from the date of shipment. This warranty extends only to the original purchaser and shall not apply to any product which, in Hart’s sole opinion, has been subject to misuse, alteration, abuse or abnormal conditions of operation or handling. Software is warranted to operate in accordance with its programmed instructions on appropriate Hart products. It is not warranted to be error free.
4 Installation Unpacking & Inspection 4 4.1 Installation Unpacking & Inspection The furnace has been carefully packed for safe shipment by traditional means. Unpacking should be done carefully. Check carefully for all parts. If any damage has occurred, you should notify the shipper immediately and make the appropriate claim. The equilibration block assembly has been packed separately in order to protect the fused silica tube from breakage during shipment.
9112A Calibration Furnace User’s Guide 4.3 “Dry-out” Period WARNING: Before initial use, after transport, and any time the instrument has not been energized for more than 10 days, the instrument needs to be energized for a “dry-out” period of 1-2 hours before it can be assumed to meet all of the safety requirements of the IEC 1010-1.
4 Installation Probe Installation shown and then insert the insulation (ceramic fiber, see MSDS in the Appendix) being careful no to bend the probe sheath. The insulation should generally block air movement in and out of the back of the fused silica tube. Install the Guard Cover to prevent physical contact with parts that become dangerously hot when in use. Be sure to connect the probes properly on the rear panel.
5 Parts and Controls Front View 5 5.1 Parts and Controls Front View Figure 1 Front View 5.1.1 Temperature Controller The temperature controller is a full PID micro-processor based instrument as indicated.
9112A Calibration Furnace User’s Guide two LED type displays. The upper display normally indicates the actual temperature while the lower display indicates the set temperature. The displays are also utilized in setup and alarm functions. Other indicators include the OP1 and OP2 indicator lights. The OP1 indicator lights when the heater is on. The OP2 is not functional on the unit. The “R” indicator lights during programmed ramping. The “M” indicator flashes if the sensor fails.
5 Parts and Controls Heater Assembly 5.2 Heater Assembly The heater is a made of fiber ceramic insulating material with imbedded heating. The heater is made up with two halves, each with a separate heating element. The heating elements are wired in parallel. The heater is primarily a radiating device and is rated for a maximum furnace operating temperature of 1100°C. Realize, however, that the higher the operating temperature, the lower the lifetime of the heater.
9112A Calibration Furnace User’s Guide Figure 3 Back View 5.2.2 Temperature Control and Cutout Sensor The temperature control sensor is a Type K Thermocouple as indicated. This sensor is 3/16 inch in diameter and 12 inches long. Its location in the block is important and can cause the gradient in the block to move back and forth. The probe is normally inserted as shown in Figures 2 and 3. The cutout sensor is the same as the control sensor, 12 inches long.
5 Parts and Controls Back View The sensor connectors are provided on the rear panel of the furnace for connecting the control and cutout thermocouples. They are Type K miniature connectors and allow for ease of system assembly and sensor replacement. 5.3 Back View See Figure 3. 5.3.1 The Power Cable The furnace is provided with a 12 gauge two conductor with ground power cable. The user must provide a connector to meet the needs of the installation. Be sure to follow electrical codes.
6 Operation Overview 6 6.1 Operation Overview The Model 9112A is basically a temperature controlled furnace utilizing a full PID micro-processor based temperature controller with a Type K thermocouple temperature sensor. The temperature controller sends a time proportional signal to the solid state relay which regulates the current to the heater. The heater power can be switched to HIGH or LOW power positions.
9112A Calibration Furnace User’s Guide CAUTION: Take care that all sensors used as references or being calibrated in the furnace are capable of withstanding the desired temperature range to be used. WARNING: Take extreme care in handling hot probes. The extreme temperatures generated in a furnace of this type can cause serious personal injury. Do not touch them on external surfaces of the furnace or set them on any other surfaces unable to withstand those temperatures. A fire hazard exists.
7 Digital Communication Interface 7 Digital Communication Interface To control the furnace through a computer, follow the instructions listed below. The program supplied is a demo program and may be altered by you for your specific needs. First make the appropriate cable assembly for your computer system. The serial communications cable attaches to the calibrator through the DB-9 connector at the back of the instrument. Figure 4 shows the pin-out of this connector and suggested cable wiring.
9112A Calibration Furnace User’s Guide • Load GWBASIC • Load and then run the program TC847.BAS • Set the Baud Rate on the Controller of the 9112A to 9600 • Set the address of your furnace to 10 or greater • Set the temperature or if needed the furnace parameters. A listing of the parameters and their meaning are in the following table. Table 2.
8 Maintenance 8 Maintenance The calibration instrument has been designed with the utmost care. Ease of operation and simplicity of maintenance have been a central theme in the product development. Therefore, with proper care the instrument should require very little maintenance. Avoid operating the instrument in an oily, wet, dirty, or dusty environment. • If the outside of the instrument becomes soiled, it may be wiped clean with a damp cloth and mild detergent.
9 Appendix A - Material Safety Data Sheets (MSDS) 9 Appendix A - Material Safety Data Sheets (MSDS) 29
MATERIAL SAFETY DATA SHEET MSDS No. M0001 Effective Date: 06/10/2003 1.
POSSIBLE CANCER HAZARD BY INHALATION. (See Section 11 for more information) CHRONIC EFFECT There has been no increased incidence of respiratory disease in studies examining occupationally exposed workers. In animal studies, long-term laboratory exposure to doses hundreds of times higher than normal occupational exposures has produced fibrosis, lung cancer, and mesothelioma in rats or hamsters. The fibers used in those studies were specially sized to maximize rodent respirability.
1986, has listed "ceramic fibers (airborne fibers of respirable size)" as a chemical known to the State of California to cause cancer. The Canadian Environmental Protection Agency (CEPA) has classified RCF as "probably carcinogenic" (Group 2).
SPILL PROCEDURES Avoid creating airborne dust. Dust suppressing cleaning methods such as wet sweeping or vacuuming should be used to clean the work area. If vacuuming, the vacuum must be equipped with a HEPA filter. Compressed air or dry sweeping should not be used for cleaning. 7. HANDLING AND STORAGE STORAGE Store in original container in a dry area. Keep container closed when not in use. HANDLING Handle ceramic fiber carefully. Limit use of power tools unless in conjunction with local exhaust.
objectives and criteria underlying each of these OEL decisions also vary. The evaluation of occupational exposure limits and determining their relative applicability to the workplace is best performed, on a case-by-case basis, by a qualified Industrial Hygienist.
control exposures to airborne fiber and the potential presence of crystalline silica. If exposure levels are known, the respiratory protection chart provided above may be applied. Potential exposure to other airborne contaminants should be evaluated by a qualified Industrial Hygienist for the selection of appropriate respiratory protection and air monitoring. Skin Protection: Wear gloves, head coverings and full body clothing as necessary to prevent skin irritation.
The University of Cincinnati is conducting an ongoing epidemiologic investigation. The evidence obtained from employees in U. S. RCF manufacturing facilities is as follows: 1) There is no evidence of any fibrotic lung disease (interstitial fibrosis) from evaluations of chest X-rays. 2) There is no evidence of an elevated incidence of lung disease among RCF manufacturing employees.
No ecological concerns have been identified. 13. DISPOSAL CONSIDERATIONS WASTE MANAGEMENT To prevent waste materials from becoming airborne during waste storage, transportation and disposal, a covered container or plastic bagging is recommended. DISPOSAL RCF, as manufactured, is not classified as a hazardous waste according to Federal regulations (40 CFR 261). Any processing, use, alteration or chemical additions to the product, as purchased, may alter the disposal requirements.
OSHA: California: Other States: pollutant. Comply with Hazard Communication Standards 29 CFR 1910.1200 and 29 CFR 1926.59 and the Respiratory Protection Standards 29 CFR 1910.134 and 29 CFR 1926.103. Ceramic fibers (airborne particles of respirable size)” is listed in Proposition 65, The Safe Drinking Water and Toxic Enforcement Act of 1986 as a chemical known to the State of California to cause cancer.
this program, please call the Unifrax Corporation Product Stewardship Information Hotline at 1-800-322-2293. On February 11, 2002, the Refractory Ceramic Fibers Coalition (RCFC) and the U.S. Occupational Safety and Health Administration (OSHA) introduced a voluntary worker protection program entitled PSP 2002, a comprehensive, multi-faceted risk management program designed to control and reduce workplace exposures to refractory ceramic fiber (RCF).
TSCA: TWA: WHMIS: Toxic Substances Control Act Time Weighted Average Workplace Hazardous Materials Information System (Canada) Revision Summary: Section 1: Added new product name. Replaces 02/11/02 MSDS. MSDS Prepared By: UNIFRAX RISK MANAGEMENT DEPARTMENT DISCLAIMER The information presented herein is presented in good faith and believed to be accurate as of the effective date of this Material Safety Data Sheet.
Thermal Ceramics MATERIAL SAFETY DATA SHEET MSDS No: 350 Date Prepared: 03/24/1992 Current Date: 7/26/2002 Last Revised: (03/20/2002) 1.
MATERIAL SAFETY DATA SHEET MSDS No: 350 Date Prepared: 03/24/1992 Current Date: 7/26/2002 Last Revised: (03/20/2002) 4. FIRST AID MEASURES RESPIRATORY TRACT (nose and throat) IRRITATION If respiratory tract irritation develops, move the person to a dust free location. See Section 8 for additional measures to reduce or eliminate exposure. EYE IRRITATION If eyes become irritated, flush immediately with large amounts of lukewarm water for at least 15 minutes.
MATERIAL SAFETY DATA SHEET MSDS No: 350 Date Prepared: 03/24/1992 Current Date: 7/26/2002 Last Revised: (03/20/2002) 8. EXPOSURE CONTROLS/PERSONAL PROTECTION MANUFACTURER'S RECOMMENDATION It is prudent to reduce exposure to respirable dusts to the lowest attainable level through the use of engineering controls such as ventilation and dust collection devices. Industrial hygiene standards and occupational exposure limits may vary between countries, state and local jurisdictions.
MATERIAL SAFETY DATA SHEET MSDS No: 350 Date Prepared: 03/24/1992 Current Date: 7/26/2002 Last Revised: (03/20/2002) 10. STABILITY AND REACTIVITY CHEMICAL STABILITY: CHEMICAL INCOMPATIBILITIES: CONDITIONS TO AVOID: HAZARDOUS DECOMPOSITION PRODUCTS: HAZARDOUS POLYMERIZATION: Stable under conditions of normal use Avoid contact with strong acids. None None Not applicable 11.
MATERIAL SAFETY DATA SHEET MSDS No: 350 Date Prepared: 03/24/1992 Current Date: 7/26/2002 Last Revised: (03/20/2002) RCRA CMS wool, as manufactured, is not classified as a hazardous waste according to Federal regulations (40 CFR 261). As manufactured, CMS wool was tested using EPA's Toxicity Characteristics Leaching Procedure (TCLP). Results showed there were no detectable contaminants or detectable leachable contaminants that exceeded the regulatory levels.
MATERIAL SAFETY DATA SHEET MSDS No: 350 Date Prepared: 03/24/1992 Current Date: 7/26/2002 Last Revised: (03/20/2002) 16. OTHER INFORMATION SUPERWOOLTM DEVITRIFICATION As produced, SuperwoolsTM are vitreous (glassy) AES Wools that do not contain crystalline silica. Continued exposure to elevated temperatures (>9000C) may cause these materials to form crystalline phases, including crystalline silica.
MATERIAL SAFETY DATA SHEET MSDS No: 350 Date Prepared: 03/24/1992 Current Date: 7/26/2002 Last Revised: (03/20/2002) ACGIH: American Conference of Governmental Industrial Hygienists ADR: Carriage of Dangerous Goods by Road (International Regulation) CAA: Clean Air Act CAS: Chemical Abstracts Service Registry Number CERCLA: Comprehensive Environmental Response, Compensation and Liability Act EPA: Environmental Protection Agency EU: European Union f/cc: Fibers per cubic centimeter HEPA: High Efficiency P