Data Sheet
3 Fluke Corporation Fluke 810 Vibration Tester
Tester specifications continued
Environmental
Operating temperature 0 °C to 50 °C (32 °F to 122 °F)
Storage temperature -20 °C to 60 °C (-4 °F to 140 °F)
Operating humidity 10 % to 95 % RH (non-condensing)
Agency approvals CHINA RoHS, CSA, CE, C TICK, WEEE
Electromagnetic compatibility EN 61326-1:2006, EN 61010:1:2001 2nd ed.
Sensor specifications
Sensor type Accelerometer
Sensitivity 100 mV/g (± 5 %, 25 °C)
Acceleration range 80 g peak
Amplitude nonlinearity 1 %
Frequency response
Z 2 - 7,000 Hz ± 3dB
X, Y 2 - 5,000 Hz ± 3dB
Power requirement (IEPE) 18 V dc to 30 V dc, 2 mA to 10 mA
Bias output voltage 12 V dc
Grounding Case grounded
Sensing element design PZT ceramic / shear
Case material 316L stainless steel
Mounting 10-32 captive socket head screw, 2-pole rare earth magnet (48 lb pull strength)
Output connector 4-Pin, M12
Mating connector M12 - F4D
Non-volatile memory TEDS 1451.4 compatible
Vibration limit 500 g peak
Shock limit 5000 g peak
Electromagnetic sensitivity, equivalent g 100 μg/gauss
Sealing Hermetic
Temperature range -50 °C to 120 °C (-58 °F to 248 °F) ± 7 %
Warranty One-year
Tachometer specifications
Dimensions (DxW) 2.86 cm x 12.19 cm (1.125 in x 4.80 in)
Weight 96 g (3.4 oz) with cable
Power Powered by 810 Vibration Tester
Detection Laser Diode Class 2
Range 6.0 to 99,999 rpm
Accuracy
6.0 to 5999.9 rpm ± 0.01 % and ± 1 digit
5999.9 to 99999 rpm ± 0.05 % and ± 1 digit
Resolution 0.1 rpm
Effective range 1 cm to 100 cm (0.4 in to 39.27 in)
Response time 1 second (> 60 rpm)