Datasheet

®
Accessories and
Ordering
Information
Included Accessories
Impact absorbing holster
Two bead probe thermocouples (54+52)
One bead probe thermocouple (51+53)
Ordering Information
FLUKE 51-II Thermometer
FLUKE 52-II Thermometer
FLUKE 53-II Thermometer
FLUKE 54-II Thermometer
Visit us on the world wide web at :
http://www.fluke.com
●●
●●
●●
●●
●●
●●
●●
The new Fluke 50 Series II contact
thermometers offer fast response and
laboratory accuracy (0.05% + 0.3°C)
in a rugged handheld test tool.
51 and 52 Series II :
Large backlit dual display shows
any combination of T1, T2 (52 only),
T1-T2 (52 only) plus MIN, MAX, or
AVG
Relative time clock on MIN, MAX,
and AVG provides a time reference
for major events
Electronic Offset function allows
compensation of thermocouple
errors to maximize overall accuracy
Measures J, K, T, and E-types of
thermocouples
Readout in °C, °F, or Kelvin (K)
Sleep mode increases battery life
Battery door allows easy battery
FeaturesFeatures
Specications
** Only the Fluke Models 53 and 54 Series II thermometers are capable of measuring N, R, and
S-type thermocouples.
Battery life: 1000 hours typical, AA
Size (HxWxD): 173 x 86 x 38 mm
Weight: 400g
Warranty: 3 years
51-II 52-II 53-II 54-II
Thermocouple types J,K,T,E J,K,T,E J,K,T,E,N,R,S J,K,T,E,N,R,S
Number of inputs Single Dual Single Dual
Time stamp Relative Time Relative Time Time of Day Time of Day
Splash/Dust resistant
Dual display with backlight
MIN/MAX/AVG recording
(T
1
-T
2
) True differential
Data logging up to 500 points
IR data port for interface
to PC
Compatible with optional
FlukeView Software
Temperature range:
J-type Thermocouples -210 °C to 1200 °C (-346 °F to 2192 °F)
K-type Thermocouples -200 °C to 1372 °C (-328 °F to 2501 °F)
T-type Thermocouples -250 °C to 400 °C (-418 °F to 752 °F)
E-type Thermocouples -150 °C to 1000 °C (-238 °F to 1832 °F)
N-type** Thermocouples -200 °C to 1300 °C (-328 °F to 2372 °F)
R** and S-type** Thermocouples 0 °C to 1767 °C (32 °F to 3212 °F)
Temperature accuracy
Above –100 °C (-148 °F) :
J, K, T, E, and N-type** ± [ 0.05% + 0.3 °C (0.5 °F) ]
R** and S-type** ± [ 0.05% + 0.4 °C (0.7 °F) ]
Below 100 °C (-148 °F) :
J, K, E, and N-types ± [ 0.20% + 0.3 °C (0.5 °F) ]
T-type ± [ 0.50% + 0.3 °C (0.5 °F) ]
replacement without breaking the
calibration seal
Optional ToolPak accessory for
hanging your thermometer
Additional features for the 53 and 54
Series II :
Data Logging up to 500 points of
data with user adjustable recording
interval
Additional thermocouple types R, S,
and N (for a total of 7 different
types)
Real time clock captures the exact
time of day when events occur
Recall function allows logged data
to be easily reviewed on the meter
display
IR communication port allows data
to be exported to optional FlukeView
®
Temperature PC software
Laboratory accuracy. Wherever you go.
Fluke 50 Series II
Thermometers