Data Sheet
3 Fluke Temperature Calibrators
Fluke 712 and 714 Temperature Calibrators
The Fluke 712 and 714 temperature calibrators deliver outstanding
performance, durability and reliability. These calibrators are compact,
lightweight and easy to carry and with a push-button interface and
are easy to use. Each calibrator is EMI tolerant, dust- and splash-
resistant and features a removable battery door for quick battery
changes.
Auto-step and auto-ramp features support remote testing.
Fluke 714 Thermocouple Calibrator
•
Measure temperature from TC probes
•
Simulate TC output
•
Operable with nine types of thermocouples
•
Calibrate linear TC transmitter with mV source function
•
Selectable °F or °C
•
Thermocouple mini-jack termination
•
Available as accessories: Fluke 700TC1 and TC2 Thermocouple
Mini-plug Kits
Fluke 712 RTD Calibrator
•
Compatible with pulsed current transmitters
•
Measure temperature from an RTD probe
•
Simulate RTD output
•
Operates with seven types of RTD
•
Measure additional RTDs using Ohms measurement function
•
Simulate additional RTDs using Ohms source function
•
°F or °C selectable
•
Four shrouded banana jacks
Functional Specifications
Range Resolution Accuracy Types
Fluke 712 Measure/simulate RTD -200 °C to 800 °C (Pt
100-385)
0.1 °C, 0.1 °F 0.2 °C, 0.4 °F
(Pt 100-385)
Pt; 100 200 500 1000
(385); Pt 100 (392); Pt 100
(392) JIS; Ni 120 (672)
Measure/simulate resistance
0 Ω to 3200 Ω
0.1 Ω .025 % + 0.1 % to
0.55 %
Fluke 714 Measure/simulate
thermocouple
–200 °C to 1800 °C,
depending on type (K,
–200 °C to 1370 °C)
0.1 °C or °F
(1 °C or °F; BRS)
0.5 °C, 0.8 °F 9 TC types; J K T E R S B
per NIST 175 and ITS-90
L U per DIN 43710 and
IPTS-68
Measure/simulate mV –10 mV to 75 mV 0.01 mV 0.025 % + 1 count
General Specifications
Maximum voltage: 30 V
Non-operating temperature: –40 °C to 60 °C
Operating temperature: –10 °C to 55 °C
Relative humidity: 95 % (10 °C to 30 °C); 75 % (30
°C to 40 °C);
45 % (40 °C to 50 °C); 35 % (50 °C to 55 °C)
Operating altitude: 3,000 m max
Shock: 1 m drop test
Vibration: Random, 2 g, 5 Hz to 500 Hz
Safety: CSA C22.2 No. 1010.1:1992 EMC: EN50082-
1:1992 and EN55022:1994 Class B
Size/weight (HxWxD): 187 mm x 87 mm x 32 mm
(7.35 in x 3.41 in x 1.25 in)
330 g (12 oz)
Size/weight (HxWxD) (with holster and Flex-
Stand™): 201 mm x 98 mm x 52 mm (7.93 in x 3.86
in x 2.06 in) 600 g (21 oz) 992 g (35 oz)
Power: 9 V battery ANSI/NEDA 1604A or IEC
6LR619V alkaline; two batteries in 718
Battery life: 4 to 20 hours, typical, depending
on functions used. Battery timeout (configurable)
extends battery life.
Warranty: Three-years
Fluke Corporation
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2008 Fluke Corporation.
Specifications subject to change without notice.
Printed in U.S.A. 8/2008 2574114 D -EN-N Rev B