Data Sheet

2 Fluke Temperature Calibrators
Measurement Accuracy
Voltage dc 30.000 V 0.02 % + 2 counts
(upper display)
20.000 V 0.02 % + 2 counts
(lower display)
100.00 mV 0.02 % + 2 counts
-10.00 mV to
75.00 mV
0.025 % + 1 count
(via TC connector)
Current dc 24.000 mA 0.02 % + 2 counts
Resistance 0.0 Ω to
400.0 Ω
0.1 Ω (4-wire)
0.15 Ω (2- and 3-wire)
401 Ω
to 1500 Ω
0.5 Ω (4-wire)
1 Ω (2- and 3-wire)
1500 Ω
to 3200 Ω
1 Ω (4-wire)
1.5 Ω (2- and 3-wire)
Source Accuracy
Voltage DC 100.00 mV 0.02 % +2 counts
10.000 V 0.02 % +2 counts
–10.00 mV to
75.00 mV
0.025 % + 1 count
(via TC connector)
Resistance 15.0 Ω
to 400.0 Ω
0.15 Ω (exc. current
0.15 mA to 0.5 mA), 0.1
Ω (exc. current 0.5 mA
to 2 mA)
401 Ω
to 1500 Ω
0.5 Ω (excitation current
0.05 mA to 0.8 mA)
1500 Ω
to 3200 Ω
1 Ω (excitation current
0.05 mA to 0.4 mA)
Specifications
Ramp functions Source functions: Voltage, current,
resistance, frequency, temperature
Ramps: Slow ramp, Fast ramp,
25 % step-ramp
Loop power function Voltage: 24 V
Accuracy: 10 %
Maximum current: 22 mA, short circuit
protected
Step functions
Source functions: voltage, resistance,
temperature
Steps: 25 % of range, 100 % of range
Environmental Specifications
Operating
temperature
–10 ºC to 55 ºC
Storage temperature –20 ºC to 71 ºC
Humidity
(Without
Condensation)
90 % 10 ºC to 30 ºC
75 % 30 ºC to 40 ºC
45 % 40 ºC to 50 ºC
35 % 50 ºC to 55 ºC
Safety Specifications
Safety rating CSA C22.2 No. 1010.1:1992
EMC EN50082-1:1992 and
EN55022:1994 Class B
RTDs and Thermocouples
Measure accuracy NI-120 0.2 ºC
PT-100 (385) 0.33 °C
PT-100 (393) 0.3 °C
PT-100 (JIS) 0.3 °C
PT-200 (385) 0.2 °C
PT-500 (385) 0.3 °C
PT-1000 (385) 0.2 °C
Resolution 0.1 °C
J 0.7 °C
K 0.8 °C
T 0.8 °C
E 0.7 °C
R 1.8 °C
S 1.5 °C
B 1.4 °C
L 0.7 °C
U 0.75 °C
N 0.9 °C
Resolution J, K, T, E, L, N, U: 0.1 °C,
0.1 °F B, R, S: 1 °C, 1 °F
XK 0.6°C
BP 1.2 °C
Source accuracy NI-120 0.2 °C
PT-100 (385) 0.33 °C
PT-100 (393) 0.3 °C
PT-100 (JIS) 0.3 °C
PT-200 (385) 0.2 °C
PT-500 (385) 0.3 °C
PT-1000 (385) 0.2 °C
Resolution 0.1 °C
Note Accuracy stated for
4-wire measurement.
J 0.7 °C
K 0.8 °C
T 0.8 °C
E 0.7 °C
R 1.4 °C
S 1.5 °C
B 1.4 °C
L 0.7 °C
U 0.75 °C
N 0.9 °C
Resolution J, K, T, E, L, N, U: 0.1 °C,
B, R, S: 1 °C
XK 0.6 °C
BP 1.2 °C
Functional specifications