Data Sheet
Fluke 707Ex Intrinsically
Safe Loop Calibrator
The fast, one-handed tool for
loop checks in Ex zones
Technical Data
Summary Specifications (18 ºC to 28 ºC, one year)
Function Range Resolution Accuracy
Voltage measure 0 to 28 V 0.001 V ± (0.015 % Rdg + 2 counts)
mA measure 0 to 24 mA 0.001 mA ± (0.015 % Rdg + 2 LSD)
mA source 0 to 24 mA
1
0.001 mA ± (0.015 % Rdg + 2 LSD)
mA simulate 0 to 24 mA
2
0.001 mA ± (0.015 % Rdg + 2 LSD)
Loop supply 24 V dc n/a 24 V ± 1 V dc, no load
Temperature Coefficient, -10 to 18 ºC, 28 to 55 ºC, ± .005 % of range per ºC;
1
Max load, 700 Ohms at 20 mA
2
Max applied voltage for simulation, 28 V
The 707Ex is an intrinsically safe loop calibra-
tor for use in explosion endangered areas. The
Fluke 707Ex is certified in accordance with the
ATEX directive (Ex II 2 G Ex ia IIC T4) in Zones
1 and 2. The 707Ex is also certified to operate
in accordance with Factory Mutual N.I. Class 1,
Division 2 areas Group A-D. The 707Ex features:
•
Large display and simple, quick-click push
rotary button for easy one-handed operation.
•
Simultaneous mA and % readout for quick,
easy, interpretation of readings.
•
mA accuracy of 0.015 %, superior to other
loop calibrators.
•
Push button with 25 % steps for fast, easy
linearity checks.
•
0-100 % “span check” for fast confirmation of
zero and span.
•
Selectable slow ramp, fast ramp, and step
ramp to provide smooth outputs for valve
slewing and loop functional tests.
•
Internal loop supply, so you can power and
read a transmitter at the same time without
carrying a DMM.
•
Measures V dc to 28 V.
•
Innovative output adjustment dial with
.001 mA and 0.1 mA resolution.
•
HART
®
compatible resistance is connected in
series with the loop supply for compatibility
with HART communicators.
ATEX and Factory
Mutual Certified