Datasheet
Test specifications
The accuracy specification for the display range is defined as ± (% reading + digit
counts) at 23 °C ± 5 °C, ≤ 75 % RH. Between 0 °C and 18 °C and between 28 °C and
40 °C, accuracy specifications may degrade by 0.1 x (accuracy specification) per °C.
The measurement range meets the service operating errors specified in EN61557-1:
1997, EN61557-2: 1997, EN61557-4: 1997, EN61557-6: 1997, DIN VDE0404-2.
Power-on test
The test indicates reversed L-N, missing PE, and
measures the mains voltage and frequency.
Display range: 90 V to 264 V
Accuracy at 50 Hz: ± (2 % + 3 counts)
Resolution: 0.1 V (1V - model 6200)
Input impedance: > 1 MΩ // 2.2 nF
Maximum input mains voltage: 264 V
Earth bond test (R
pe
)
Display range: 0 to 19.99 Ω
Accuracy (after Bond Test zeroing):
± (2.5 % + 4 counts)
Resolution: 0.01 Ω
Test current: 200 mA ac -0 % +40 % into 1.99 Ω
25 A ac ± 20 % into 25 mΩ at 230 V
Open circuit voltage: > 4 V ac, < 24 V ac
Bond test zeroing: Can subtract up to 1.99 Ω
Insulation test (R
ISO
)
Display range: 0 to 299 MΩ
Accuracy: ± (5 % + 2 counts) from 0.1 MΩ to 50 MΩ
± (10 % + 2 counts) from 50 MΩ to 299 MΩ
Resolution: 0.01 MΩ (0 to 19.99 MΩ)
0.1 MΩ (20 MΩ to 199.9 MΩ)
1 MΩ (200 MΩ to 299 MΩ)
Test voltage: 500 V dc –0 % +25 % at 500 kΩ load or
(6500-2 only) 250 V dc -0 % +25 % at 250 kΩ load
Test current: >1 mA at 500 kΩ load, < 15 mA at 0Ω
Auto discharge time: < 0.5 s for 1 μF
Max. capacitive load: Operational up to 1 μF
Touch current test
Display range: 0 to 1.99 mA ac
Accuracy: ± (4 % + 2 counts)
Resolution: 0.01 mA
Internal resistance (via probe): 2 kΩ
Measuring method: Probe
The appliance under test is energized
at mains potential.
Substitute leakage current
Display range: 0 to 19.99 mA ac
Accuracy: ± (2.5 % + 3 counts)
Resolution: 0.01 mA
Test voltage: 100 V ac ± 20 %
Operational error: 10 %
RCD Test: trip current
(6500-2 only)
Operational error: ± 10 %
Nominal: 30 mA
Accuracy: ± 5 %
RCD Test: trip time
(6500-2 only)
Standard requirement:
61557 Part 6; tolerance of rated test
current 0 % to +10 %
Operational error: ± 10 %
RCD type: AC general purpose 30 mA
Display range: 310 ms
Resolution: 0.1 ms
Accuracy: 3 ms
Trip time limit at 100 % (30 mA): 300 ms
Trip time limit at 500 % (150 mA): 0 ms
Load/Leakage test: load current
UK Display range: 0 to 13 A
Accuracy: ± (4 % + 2 counts)
Resolution: 0.1 A
The appliance under test is energized at mains potential.
AU Display range: 0 to 10 A
Accuracy: ± (4 % +2 counts)
Resolution: 0.1 A
The appliance under test is energized at mains potential.
Load/Leakage test: load power display
range 230 V mains
UK: 0.0 VA to 3.2 kVA
AU: 0.0 VA to 2.4 kVA
Accuracy: ± (5 % + 3 counts)
Resolution: 1 VA (0 to 999 VA), 0.1 kVA (>1.0 kVA)
The appliance under test is energized at mains potential.
Load/Leakage test: Leakage current
Display range: 0.25 mA to 19.99 mA
Accuracy: ± (4 % + 4 counts)
Resolution: 0.01 mA
The appliance under test is energized at mains potential.
PELV test
Accuracy at 50 Hz: ± ( 2 % + 3 counts)
Overload protection: 300 V rms
Warning threshold: 25 V rms