715 English Instruction Sheet Page 1 ® 715 Volt/mA Calibrator Instruction Sheet Introduction The Fluke 715 Volt/mA Calibrator is a source and measurement tool for 0 to 24 mA current loop testing and dc voltage from 0 to 20/25 V. The calibrator does not source and measure simultaneously. Your calibrator is supplied with a holster, a set of test leads, an installed 9 V alkaline battery, and this instruction sheet. Summary of Calibrator Capabilities Function Range Resolution dc mV input 0 to 200 mV 0.
715 English Instruction Sheet Page 2 If the calibrator is damaged or something is missing, contact the place of purchase immediately. Contact your Fluke distributor for information about accessories. To order replacement parts or spares, see “Replacement Parts.
715 English Instruction Sheet Page 3 Safety Information Use the calibrator only as specified in this sheet, otherwise the protection provided by the calibrator may be impaired. A Warning identifies conditions and actions that pose hazard(s) to the user; a Caution identifies conditions and actions that may damage the calibrator or the equipment under test.
715 English Instruction Sheet Page 4 1. With the Calibrator OFF, press the O key. P.S.xx is displayed, where xx is the turn-off time in minutes. OFF means the power saver is disabled. 2. Press ] and/or } to increase or decrease the turnoff time in minutes. 3. To disable, press } until the display shows OFF. HART TM Resistor Mode The calibrator has a user selectable 250 ohm HARTTM resistor to facilitate use with HARTTM communication devices.
715 English Instruction Sheet Page 5 Measuring dc Volts INPUT mV or V depending on range V mV mA % V mV Press so that INPUT is on the display Press to toggle mV and V ranges it01i.eps Sourcing dc Volts OUTPUT mV or V depending on range V mV Press so that OUTPUT is on the display mA % V mV Press to toggle mV and V ranges it02i.
715 English Instruction Sheet Page 6 Measuring dc mA INPUT mA % V mV mA % Press to toggle mA and % of 4-20 mA scale Press so that INPUT is on the display it03i.
15 English Instruction Sheet Page 7 Measuring dc mA with Loop Power INPUT % mA % V mV mA % Press to toggle mA and % of 4-20 mA scale Press so that INPUT is on the display +24V output Two-wire transmitter HARTTM Communicator it06i.
715 English Instruction Sheet Page 8 Using the Current Output Modes The calibrator provides current output in mA or percent display. Percent is -25.00 to 125.00%, where 0% is 4 mA, and 100% is 20 mA. In source mode, the calibrator supplies the current. In simulate mode, the calibrator simulates a two-wire transmitter in an externally-powered current loop. Sourcing mA Use source mode whenever you need to supply current into a passive circuit such as a current loop with no loop supply.
715 English Instruction Sheet Page 9 Simulating a Transmitter Use simulate mode when an external 24 to 30 V loop power supply is available. Insert the test leads into the mA SIMULATE − and + jacks as shown below. OUTPUT Press to step up/down 4 mA (25%) Press so that OUTPUT is on the display mA % V mV Press to toggle mA and % of 4-20 mA scale mA % Press to scroll up/down 0.001 mA. Hold down to scroll faster. COM 30V dc max. Power supply it05i.
715 English Instruction Sheet Page 10 Maintenance For maintenance procedures not described in this sheet, contact a Fluke Service Center. In Case of Difficulty • Check the battery and test leads. Replace as necessary. • Review this sheet to make sure you are using the correct jacks and pushbuttons. If the calibrator needs repair, contact a Fluke Service Center. If the calibrator is under warranty, see the warranty statement below for terms.
715 English Instruction Sheet Page 11 Cleaning Periodically wipe the case with a damp cloth and detergent; do not use abrasives or solvents. Calibration Calibrate your calibrator once a year to ensure that it performs according to its specifications. A calibration manual is available (PN 686540). Call 1-800-526-4731 from the USA and Canada. In other countries, contact a Fluke Service Center.
715 English Instruction Sheet Page 12 Replacement Parts and Accessories Replacement Parts Item Description PN or Model no. Qty.
715 English Instruction Sheet Page 13 MP85 S1 V V m A m % MP8 BT1 MP86 H2, 3, 4 MP89, 90 H5, 6 MP92 Instruction Sheet AC72 Alligator Clips TL75 Test Lead Set Holster it10c.
715 English Instruction Sheet Page 14 Specifications Specifications are based on a one year calibration cycle and apply from +18°C to +28°C unless stated otherwise. “Counts” means number of increments or decrements of the least significant digit. DC V Input and Output Range 200 mV Resolution Accuracy, ±(% of Reading + Counts) 0.01 mV 0.015% + 2 0.001 V 0.
715 English Instruction Sheet Page 15 General Specifications Maximum voltage applied between any jack and earth ground or between any two jacks: 30 V Storage temperature: -40°C to 60°C Operating temperature: -10°C to 55°C Operating altitude: 3000 meters maximum Temperature coefficient: ±0.
715 English Instruction Sheet Page 16 LIMITED WARRANTY & LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for three years from the date of purchase. This warranty does not cover fuses, disposable batteries or damage from accident, neglect, misuse or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Fluke’s behalf.
Supplement Title: 715 Inst.Sht. Supplement Issue: 6 Part Number: 650314 Issue Date: 1/14 Print Date: July 1997 Page Count: 2 Revision/Date: 3, 8/05 This supplement contains information necessary to ensure the accuracy of the document described above. © 2007-2014 Fluke Corporation. All rights reserved.
715 Instruction Sheet Supplement Change #1 Under General Specifications, add the following: Protection Class: Pollution Degree II Change #2 Under Specifications, table DC V Input and Output change the following: From: Voltage drive capability: 1 mA To: Voltage drive capability: SN < 9494143 = 1 mA SN > 9494143 = 10 mA 10 V, 3 mA > 10 V Change #3 On the second panel, under International Symbols, add the following: Conforms to Australian C-Tick directives Change #4, 47219 Under Specifications, D
Supplement 714 Instruction Sheet Change #6, 67381 On the second panel, under International Symbols, add: Conforms to relevant South Korean EMC Standards. Under General Specifications, add: Electromagnetic Compatibility ......Applies to use in Korea only. Class A Equipment (Industrial Broadcasting & Communication [1] Equipment) [1] This product meets requirements for industrial (Class A) electromagnetic wave equipment and the seller or user should take notice of it.