Datasheet
8 Fluke Corporation Fluke MDA-510 and MDA-550 Motor Drive Analyzer
Specifications
Measurement Function Specification
DC voltage (V dc)
Maximum voltage with 10:1 or 100:1
probe
1000 V
Maximum resolution with 10:1 or 100:1
probe
1 mV
Full scale reading 999 counts
Accuracy at 4 s to 10 us/div ± (3 % + 6 counts)
AC voltage (V ac)
Maximum voltage with 10:1 or 100:1
probe
1000 V
Maximum resolution with 10:1 or 100:1
probe
1 mV
Full scale reading 999 counts
50 Hz ± (3 % + 10 counts) - 0.6 %
60 Hz ± (3 % + 10 counts) - 0.4 %
60 Hz to 20 kHz ± (4 % + 15 counts)
20 kHz to 1 MHz ± (6 % + 20 counts)
1 MHz to 25 MHz ± (10 % + 20 counts)
True-rms voltage (V ac+dc)
Maximum voltage with 10:1 or 100:1
probe
1000 V
Maximum resolution with 10:1 or 100:1
probe
1 mV
Full scale reading 110 0 c ou nt s
DC to 60 Hz ± (3 % + 10 counts)
60 Hz to 20 kHz ± ( 4 % + 15 counts)
20 kHz to 1 MHz ± (6 % + 20 counts)
1 MHz to 25 MHz ± ( 10 % + 20 counts)
PWM voltage (V pwm)
Purpose To measure on pulse width modulated signals, like motor drive inverter outputs
Principle Readings show the effective voltage based on the average value of samples over a whole number
of periods of the fundamental frequency
Accuracy As Vac+dc for sinewave signals
Peak voltage (V peak)
Modes Max peak, min peak, or pk-to-pk
Maximum voltage with 10:1 or 100:1
probe
1000 V
Maximum resolution with 10:1 or 100:1
probe
10 mV
Accuracy
Max peak, min peak ± 0.2 division
Pk-to-pk ± 0.4 division
Full scale reading 800 counts