User's Manual

Fluke 434-II/435-II/437-II
Handleiding
27-6
Volt/Amps/Hertz Meetbereik Resolutie Nauwkeurigheid
Hz
2
Fluke 435-II /437-II @ 50
Hz nominaal
Fluke 435-II /437-II @ 60
Hz nominaal
Fluke 437-II @ 400 Hz
nominaal
Fluke 434-II @ 50Hz
nominaal
Fluke 434-II @ 60Hz
nominaal
42.5 ... 57.5 Hz
51 ... 69 Hz
340 ... 460 Hz
42.5 ... 57.5 Hz
51 ... 69 Hz
0.001 Hz
0.001 Hz
0.1 Hz
0.001 Hz
0.001 Hz
± 0.001 Hz
± 0.001 Hz
± 0.1 Hz
± 0.01 Hz
± 0.01 Hz
Power Meetbereik Resolutie Nauwkeurigheid
Watt (VA, var)
i430flex-TF
1 mV/A
max 6000 MW
max 2000 MW
0.1 W … 1 MW
0.1 W … 1 MW
± 1% ± 10 counts
± 1% ± 10 counts
Power Factor (Cos ϕ /
DPF)
0…1 0.001 ± 0.1% @ nominale belasting-
condities
Energie Meetbereik Resolutie Nauwkeurigheid
kWh (kVAh, kvarh)
i430flex-TF 10x
Hangt af van stroomtang scaling en
V nominaal
± 1% ± 10 counts
Energieverlies
i430flex-TF 10x
Hangt af van stroomtang scaling en
V nominaal
± 1% ± 10 counts
Exclusief bedradingsweerstand
nauwkeurigheid
Harmonischen Meetbereik Resolutie Nauwkeurigheid
Harmonic order (n) DC, 1..50 Groepering: Harmonischen groepen volgens IEC 61000-4-7
Inter-Harmonic order Off, 1..50 Groepering: Harmonischen en Interharmonischen-subgroepen
volgens IEC 61000-4-7
Volts %f 0.0 … 100.0% 0.1% ± 0.1% ± n x 0.1%
Volts %r 0.0 … 100.0% 0.1% ± 0.1% ± n x 0.4%
Volts Absolute 0.0 … 1000 V 0.1 V ± 5% (*)
Volts THD 0.0 … 100.0% 0.1% ± 2.5%
Amps %f 0.0 … 100.0% 0.1% ± 0.1% ± n x 0.1%
Amps %r 0.0 … 100.0% 0.1% ± 0.1% ± n x 0.4%
Amps Absolute 0.0 … 600 A 0.1 A ± 5% ± 5 counts
Amps THD 0.0 … 100.0% 0.1% ± 2.5%
Watts %f of %r 0.0 … 100.0% 0.1% ± n x 2%
Watts Absolute Afhankelijk van stroomtangschaal en V nom ± 5% ± n x 2% ± 10 counts