User's Manual

Fig. 1. PM9080
Fig. 2. Adapter DB-9 to Micro USB
Fig. 3. PM8907 Fig. 4. GPS 430
Using the GPS 430
1. Connect the PM8907 power adapter output to the GPS 430 power input.
2. Check the preselected voltage setting on the PM8907/808, see the PM8907/808
Instructions below.
3. Connect the PM8907 to the line voltage.
4. Connect the PM9080 Optical Interface Cable to the GPS 430 and to the Power Quality
Analyzer.
5. Verify that the GPS 430 data are received by the Power Quality Analyzer:
Fluke 433/434/435: Verify that the GPS 430 data are received by the Power Quality
Analyzer:
select SETUP, use the up/down arrow keys to hightlight Time, ENTER, function key F2 -
GPS ON to start the GPS TEST. See also the Power Quality Analyzer Users Manual.
Fluke 434-II/435-II/437-II (Series II): press in sequence the SETUP key, press function
key F4 – MANUAL SETUP, use the up/down arrow keys to hightlight Time, press
ENTER, press F2 – GPS ON to start the GPS TEST. See also the Power Quality
Analyzer Users Manual.
Note: For proper reception, the GPS antenna should not be covered by metal.
PM8907/808 Instructions
Please read these instructions carefully before using the Power Adapter.
Caution
Before you connect the PM8907/808 to the local line, first check the preselected voltage
setting on this unit. You can find the voltage selector switch on the bottom of the
PM8907/808. If necessary, select the corresponding line voltage with the slide switch
and ensure that the proper voltage is visible in the window of the slide selector. (see
Fig. 2.)
For connection to the mains outlet, use a power cord that complies with national
standards.
Observe the polarity 5-mm power jack; the outer contact is plus (+) polarity.
Do not open the PM8907, it is not serviceable.
PM8907/808 Specifications
Universal switchable adapter 115V ±10% or 230V ±15%, with plug EN60320-2.2G
Line frequency: 50, 60, and 400 Hz ±10%.
Output voltage: 15V.
Output current: 300 mA.
Power consumption: 7W maximum.
Output plug: 5-mm power jack per DIN 45232.
Plus (+) voltage connected to outer contact of the 5-mm power jack.
Symbols
International electrical symbols used are explained below.
See explanation on Instruction Sheet.
DOUBLE INSULATION (Protection Class).
Conforms to relevant European standards.
Do not dispose of this product as unsorted municipal waste. Go to
Fluke's website for recycling information.