Meter inputs Via 4 mm banana inputs, fully isolated from
scope inputs and scope ground
Via BNC scope inputs
Number of readings One at a time Up to 4 simultaneously
Maximum resolution 5,000 counts 999 counts
Input impedance 1 MΩ ± 1 % // 14 pF ± 2 pF
Advanced meter functions Auto/manual ranging, relative measurements (Zero reference), TrendPlot
recording
The specified accuracy is valid over the temperature range 18 °C to 28 °C
Add 10 % of specified accuracy for each degree C below 18 °C or above 28 °C
Voltage
Vdc accuracy ± (0.5 % + 5 counts) ± (1.5 % + 5 counts)
Vac true rms accuracy
15 Hz to 60 Hz: ± (1 % + 10 counts) ± (1.5 % + 10 counts)
60 Hz to 1 kHz: ± (2.5 % + 15 counts)
60 Hz to 20 kHz: ± (2.5 % + 15 counts)
Vac+dc true rms accuracy
15 Hz to 60 Hz: ± (1 % + 10 counts) ± (1.5 % + 10 counts)
60 Hz to 1 kHz: ± (2.5 % + 15 counts)
60 Hz to 20 kHz: ± (2.5 % + 15 counts)
Voltmeter ranges 500 mV, 5 V, 50 V, 500 V, 1,000 V
Resistance
Ranges 500 Ω, 5 kΩ, 50 kΩ, 500 kΩ, 5 MΩ, 30 MΩ
Accuracy ± (0.6 % + 5 counts)
Other meter functions
Continuity Beeper on < 50 Ω (± 30 Ω)
Diode test Up to 2.8 V
Current (A) Adc, Aac, Aac+dc using an optional current clamp or shunt
Scaling factors: 0.1 mV/A, 1 mV/A to 100 V/A and 400 mV/A
Temperature With optional accessories. Scale factors 1 °C/mV or 1 °F/mV
Meter Modes