Data Sheet
ScopeMeter 190 Series II Fluke Corporation 5
190-062 190-102 190-202 190-104 190-204
ScopeRecord
™
Roll Mode
Dual or multiple input waveform storage mode, using deep memory
Source and display Input A, Input B, Dual
All channels sampled simultaneously
Any combination of inputs,
up to 4 channels
All channels sampled simultaneously
Bandwidth 20 MHz or 20 kHz, user selectable
Memory depth 30,000 data points, each holding min/max pair of information
Min/max values Min/max values are created at samples that are measured at high sample rate
ensuring capture and display of glitches
Recording modes Single sweep, continuous roll,
Start-on-Trigger (through external),
Stop-on-Trigger (through external)
Single sweep, continuous roll,
Start-on-Trigger (through any channel),
Stop-on-Trigger (through any channel)
Stop-on-trigger ScopeRecord mode can be stopped by an individual trigger event, or by an interruption of a repetitive
trigger signal, through any input channel (through External on 190-XX2 Series)
Horizontal scale Time from start, time of day
Zoom Ranges from full record overview to zoom in up to sample level, at any record length
Memory Two multiple input ScopeRecord waveforms can be saved internally for later recall and analysis
Direct storage on external flash memory drive through USB host port
ScopeRecord
™
Roll mode sample rate and recording timespan
Time base range 5 ms/div ~ 2 min/div
Recorded timespan 6 sec ~ 48 hr
Time/division in ‘view all’ mode 0.5 s/div ~ 4 h/div
Glitch capture 8 ns
Sample rate 125 MS/s
Resolution 200 μsec ~ 4.8 sec
Trendplot
™
Recording
Multiple channel electronic paperless recorder
Graphically plots, displays and stores results of up to four automatic scope measurements or a DMM-reading over time
Source and display Any combination of scope measurements, made on any of the input channels, or DMM reading
(2-channel instruments)
Memory depth 18,000 points (sets) per measurement
Each recorded sample point contains a minimum, a maximum and an average value,
plus a date- and timestamp
Ranges Normal view: 5 s/div to 30 min/div
In view-all mode: 5 min/div to 48 hr/div (overview of total record)
Recorded time span Up to 22 days, with a resolution of 102 seconds
Recording mode Continuous recording, starting at 5 s/div with automatic record compression
Measurement speed 3 automatic measurements per second or more
Horizontal scale Time from start, time of day
Zoom Up to 64x zoom-out for full record overview, up to 10x zoom-in for maximum detail
Memory Two multiple input TrendPlot records can be saved internally for later recall and analysis
Direct storage on external flash memory drive through USB host port
Cursor measurements – all recorder modes
Source Any waveform trace in any waveform display mode (Scope, ScopeRecord or TrendPlot)
Dual vertical lines Cursors may be used to identify Min, Max or Average value of any datapoint in a record, with time
between cursors, time from start or absolute time
Recorder Modes