Safety Information: Read First
5
Warning
To avoid electrical shock or fire if a product
input is connected to more than 42 V peak
(30 Vrms) or 60 V dc:
Use only insulated voltage probes, test leads
and adapters supplied with the product, or
indicated by Fluke as suitable for the Fluke
190-104 and 190-204 ScopeMeter series.
Before use, inspect voltage probes, test leads
and accessories for mechanical damage and
replace when damaged.
Remove all probes, test leads and
accessories that are not in use.
Always connect the power adapter first to the
ac outlet before connecting it to the product.
Do not touch voltages >30 V ac rms, 42 V ac
peak, or 60 V dc.
Do not connect the ground spring (figure 1,
item d) to voltages higher than 42 V peak
(30 Vrms) from earth ground.
Do not apply more than the rated voltage,
between the terminals or between each
terminal and earth ground.
Do not apply input voltages above the rating
of the instrument. Use caution when using
1:1 test leads because the probe tip voltage
will be directly transmitted to the product.
Do not use exposed metal BNC connectors.
Fluke offers cables with plastic, safety
designed BNC connectors suitable for the
ScopeMeter product, see Chapter 7 ‘Optional
accessories’.
Do not insert metal objects into connectors.
Use the product only as specified, or the
protection supplied by the product can be
compromised.
Carefully read all instructions.
Do not use the product if it operates
incorrectly.
Do not use and disable the product if it is
damaged.
Keep fingers behind the finger guards on the
probes.
Use only correct measurement category
(CAT), voltage, and amperage rated probes,
test leads, and adapters for the measurement.