User manual

Fluke 190-104, Fluke190-204
Users Manual
102
Edge Trigger
Screen Update............Free Run, On Trigger, Single Shot
Source .............................................................. A, B, C, D
Slope .......................................... Positive, Negative, Dual
Trigger Level Control Range.......................... ±4 divisions
Trigger Sensitivity
DC to 5 MHz at >5 mV/div ......................... 0.5 divisions
DC to 5 MHz at 2 mV/div and 5 mV/div........... 1 division
200 MHz (FLUKE 190-204)............................. 1 division
250 MHz (FLUKE 190-204)........................... 2 divisions
100 MHz (FLUKE 190-104)............................. 1 division
150 MHz (FLUKE 190-104)........................... 2 divisions
Video Trigger
Standards ....PAL, PAL+, NTSC, SECAM, Non-interlaced
Modes...................... Lines, Line Select, Field 1 or Field 2
Source ............................................................................A
Polarity................................................. Positive, Negative
Sensitivity........................................0.7 division sync level
Pulse Width Trigger
Screen Update.............................On Trigger, Single Shot
Trigger Conditions ...........<T, >T, =T (±10 %), T(±10 %)
Source ............................................................................A
Polarity ..................................... Positive or negative pulse
Pulse Time Adjustment Range...........0.01 div. to 655 div.
with a minimum of 300 ns (<T, >T) or 500 ns (=T, T),
a maximum of 10 s,
and a resolution of 0.01 div. with a minimum of 50 ns
Continuous Auto Set
Autoranging attenuators and time base, automatic
Connect-and-View™ triggering with automatic source
selection.
Modes
Normal.....................................15 Hz to max. bandwidth
Low Frequency......................... 1 Hz to max. bandwidth
Minimum Amplitude A, B, C, D
DC to 1 MHz........................................................ 10 mV
1 MHz to max. bandwidth.................................... 20 mV
Automatic Capturing Scope Screens
Capacity .............................................100 Scope Screens
For viewing screens, see Replay function.