User manual

Specifications
Four Input Oscilloscope
8
101
Table 2. Record Lenth (Samples/points per input)
Mode Glitch Detect On Glitch Detect Off Max. Samplerate
Scope - Normal 300 min/max pairs 3k true samples compressed
into 1 screen (300 samples
per screen)
Scope - Fast 300 min/max pairs 300 true samples
Scope - Full 300 min/max pairs 10k true samples,
compressed into 1 screen.
Use Zoom and Scroll to see
waveform details
2.5 GS/s (FLUKE 190-204
and 1 or 2 channels on)
1.25 GS/s (All models and 3
or 4 channels on)
Scope Record Roll 30k to 37.5k samples 4x 125 MS/s
Trend Plot > 18k min/max/ average
values per measurement
Up to 5 measurements per
second
Glitch Detection
5 μs to 120 s/div........... displays glitches as fast as 8 ns
Waveform Display............................................A, B, C, D,
Math (+, -, x, X-Y mode, spectrum)
Normal, Average, Persistence, Reference
Time Base Accuracy ...................... ±(100 ppm + 0.04 div)
Trigger and Delay
Trigger Modes........................................Automatic, Edge,
Video, Pulse Width, N-Cycle
Trigger Delay.................................. up to +1200 divisions
Pre Trigger View..............................one full screen length
Delay ................................................. -12 div to +1200 div
Max. Delay ................................................. 60 s at 5 s/div
Automatic Connect-and-View Trigger
Source...............................................................A, B, C, D
Slope ...........................................Positive, Negative, Dual