User's Manual

772/773
Calibration Manual
14
mA DC Source Adjust Procedure
To adjust the mA Source function, use the Calibrator to apply the necessary Meter input
and do the steps in Table 9.
Table 9. mA Source Adjustment Procedure
Step Meter LCD display Action
1 4.00 mA Adjust until Meter output is 4.00 mA, push H
2 20.00 mA Adjust until Meter output is 20.00 mA, push H
4 Save Push H
Volts DC Source Adjust Procedure (773 Only)
To adjust the Volt Source function, use the Calibrator to apply the necessary Meter input
and do the steps in Table 10.
Table 10. Volt Source Adjustment Procedure
Step Meter LCD display Action
1 0.00 V Adjust until Meter output is 0.00 V, push H
2 10.00 V Adjust until Meter output is 10.00 V, push H
4 Save Push H