Once the screen in Figure 4 appears:
1. Press M twice.
2. Press E twice.
The calibration mode screen appears, see Figure 5.
fcs004.bmp
Figure 5. Calibration Mode Screen
Use the procedure in the next section to adjust the Clamp Meter.