Change #2, 63545
On page 4, add the following to the Symbols table:
Conforms to relevant South Korean EMC Standards.
Conforms to CAN/CSA-C22.2 No. 61010-1, second
edition, including Amendment 1.
Change #3, 65704
On page 9, under General Data:
Change: Battery Eliminator BE345
Input……110V / 230V, 50/60 Hz
Output….15 V dc, 300 mA
To: Battery Eliminator GFP151DA-120125B-1
Input……100 V to 240 V, 50/60 Hz, 0.36 A
Output….12 V dc, 1.25 A
On page 10, in the Note:
Change: BE345
To: GFP151DA-120125B-1
On page 10, under the Warning remove the second bullet and in bullets one and three:
Change: BE345
To: GFP151DA-120125B-1
On page 10, remove Figure 2.
On page 12, under Battery and Battery Eliminator Check in step 1:
Change: BE345
To: GFP151DA-120125B-1
On page 15, Table 2:
Change: BE345
To: GFP151DA-120125B-1
On page 28, Table 4 in the part number column:
Change: BE345
To: GFP151DA-120125B-1
Change #4, 66326
On page 4, Table 1, replace CAT III and CAT IV, with:
CAT III
Measurement Category III is applicable to test and measuring circuits connected to
the distribution part of the building’s low-voltage MAINS installation.
CAT IV
Measurement Category IV is applicable to test and measuring circuits connected at
the source of the building’s low-voltage MAINS installation.