User's Manual
  Digital Multimeter 
  Calibration Adjustments 
 15 
Table 5. Accuracy Tests (cont.) 
Step 
Test 
Function 
Range 5520A Output 
UUT Measurement 
Low Limit  High Limit 
42   
DC Microamps 
600 μA 330 μA  328.9 μA 331.1 μA 
43 6000 μA 3300 μA 3291 μA 3309 μA 
44   
Capacitance 
10 nf  Open input 
[1]
  0.28 nf  0.48 nf 
45  100 nf  5 nf 
[2]
  4.7 nf  5.3 nf 
46 100 μf 9.5 μf  9.2 nf  9.8 nf 
47 
 
Low Pass 
Filter 
1000 V  400 V, 400 Hz  372 V  408 V 
48  1000 V  400 V, 800 Hz 
[3]
  226 V  340 V 
49  VDC Peak 
Min/Max 
6 V  8 Vpp, 500 Hz Sq. Wave, 
DC offset 2 V 
-1.898 V to -
2.102 V 
5.896 V to 
6.104 V 
50 mVdc 
Temperature
 [4]
0 °C -1.0 °C 1.0 °C 
51 
100 °C 98.0 °C 102.0 °C 
[1]  Remove test leads from unit. 
[2]  Use REL to compensate for internal Meter and lead capacitance (must disconnect test leads from calibrator before pushing REL) 
[3]  The Meter accuracy is not specified at this input signal frequency with Low-pass filter selected. The display reading shown, checks 
that the Low-pass filter is active and follows an expected roll-off curve. 
[4]  To ensure accurate measurement, the Meter and thermocouple adapter must be at the same temperature. After connecting the 
thermocouple adapter to the Meter allow for reading to stabilize before recording display reading. 
Backlight Functional Test 
A backlight test is done to make sure the backlight comes on with the first push of H. 
The second push causes the backlight to be brighter and a third push turns off the 
backlight. 
Calibration Adjustments 
If one or more of the accuracy tests shows a measurement that is not between the high 
and low limits, you can do adjustments. This adjustment procedure sets the Meter to 
operate to its specifications. 
Note 
If the adjustment routine is stopped before you complete the procedure, no 
changes are made to the calibration constants in memory. 
You must set the Meter to the CAL mode and type in a password to do calibration 
adjustments. 
Cal Mode Button Functions 
For the CAL mode, some of the buttons on the Meter have alternative functions. Some 
buttons function differently at different points in the calibration adjustment procedure. 
Table 6 is a list of the Meter buttons and their function for the calibration adjustment 
procedure. 