User's Manual
28 II EX Getting Started  Manual Supplement 
 2 2/12 
Ex-Certification Data 
  Ex-Type certificate no: PTB 11 ATEX 2028 X 
   IECEx PTB 11.0080X 
  Ex-Designation: ATEX: II 2G Ex ia IIC T4 Gb 
   II 2D Ex ia IIIC T130 C Db 
   I M1 Ex ia I Ma 
IECEx: Ex ia IIC T4 Gb 
 Ex ia IIIC T130 C Db 
 Ex ia I Ma 
 CE: CE0102 
 Operating Temperature: -15 C to 50 C 
  Storage Temperature: -40 C to +60 C 
  Batteries: 3 AAA Alkaline batteries, NEDA 24A IEC 
LR03. Table 7 shows the approved batteries for this 
Product. 
For connections to intrinsically-safe circuits, observe 
these Product connections: 
Voltage – measurement input U
i
 = 65 V: 
U
0
 = 9.54 V   
C
i
 = negligible 
I
0 
= 3.7 mA  I
i
 = negligible 
L
i
 = negligible 
P
0 
= negligible  Ri = 2.47K 
Lo/Co 
Lo/mH 1000 100  2  0.5  0.1  0.01 
Co/F 0  0.61 1  1.4  2.1  3.6 
Current – measurement input I
i
 = 5 A: 
U
0
 = 0 V   U
i
 = 65 V 
C
0 
= 1000 F C
i
 = negligible 
I
0 
= 0mA   
L
0 
= 1000 mH  L
i
 = negligible 
P
0 
= 0 mW 
mA/A Jack 
U
0
 = 1.95 V  U
i
 = 65 V 
C
i
 = negligible 
I
0 
= 9.7 A I
i
  = Internally limited by a 440 mA fuse 
L
i
 = negligible 
P
0 
= negligible 