21
Table 5. Fluke 27 II Calibration Adjustment Steps (cont.)
Function
(Switch Position)
Adjustment Step Input Value
C-36 0.0 μA ac, 0 Hz
C-37 30.0 μA, 60 Hz
C-38 60.0 μA. 60 Hz
C-39 600.0 μA, 60 Hz
C-40 6000.0 μA, 60 Hz
C-41 600.0 μA, 0 Hz
(Microamps)
C-42 6000.0 μA, 0 Hz
Table 6. Fluke 28 II Calibration Adjustment Steps
Function
(Switch Position)
Adjustment Step Input Value
C-01 600.0 mV, 60 Hz
C-02 600.0 mV, 20 kHz
C-03 6.000 V, 60 Hz
C-04 6.000 V, 20 kHz
C-05 60.00 V, 60 Hz
C-06 60.00 V, 20 kHz
C-07 600.0 V, 60 Hz
(AC Volts)
C-08 600.0 V, 10 kHz
C-09 6.000 V
C-10 60.00 V
(DC Volts)
C-11 600.0 V
C-12 600.0 mV
(DC Millivolts)
C-13 60.00 mV
C-14 600.0 Ω
C-15 6.000 kΩ
C-16 60.00 kΩ
C-17 600.0 kΩ
C-18 6.000 MΩ
C-19 0.000 Ω
(Ohms)
C-20 50.0 MΩ
(Diode Test)
C-21 3.000 V
C-22 6.000 A, 60 Hz
(Amps)
C-23 6.000 A dc
C-24 60.00 mA, 60 Hz
C-25 400.0 mA, 60 Hz
C-26 60.00 mA dc
(Amps)
C-27 400.0 mA dc