Data Sheet
DETAILED SPECIFICATIONS
Fluke 114, 115, 116 and 117
Digital Multimeters
Accuracy specifications (all models)
Function Range Resolution Accuracy ± ([% of Reading]+ [Counts]) Model
DC millivolts
600.0 mV
0.1 mV
0.5 % + 2 114, 115, 116, 117
DC volts 6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
0.5 % + 2 114, 115, 116, 117
DC, 45 Hz to 500 Hz 500 Hz to 1 kHz
Auto-V LoZ true-rms
(114, 116, 117 on ly )
600.0 V
0.1 V 2.0 % + 3 4.0 % + 3
114, 116, 117
45 Hz to 500 Hz 500 Hz to 1 kHz
AC millivolts true-rms
600.0 V 0.1 V 1.0 % + 3 2.0 % + 3 114, 115, 116, 117
AC volts true-rms 6.000 V
60.00 V
600.0 V
0.001 V
0.01 V
0.1 V
1.0 % + 3 2.0 % + 3 114, 115, 116, 117
Continuity (115, 116 only)
600 1
Beeper on < 20 , off > 250 ; detects
opens or shorts of 500 s or longer
114, 115, 116, 117
Ohms 600.0
6.000 k
60.00 k
600.0 k
6.000 M
40.00 M
0.1
0.001 k
0.01 k
0.1 k
0.001 M
0.01 M
0.9 % + 2
0.9 % + 1
0.9 % + 1
0.9 % + 1
0.9 % + 1
5.0 % + 2
114, 115, 116, 117
Diode test
2.000 V
0.001 V 0.9 % + 2
115, 116 , 117
Capacitance (115, 116, 117 only) 1000 nF
10.00 F
100.0 F
9999 F
-
1 nF
0.01 F
0.1 F
1 F
-
1.9 % + 2
1.9 % + 2
1.9 % + 2
100 F to 1000 F: 1.9 % + 2
> 1000 F: 5 % + 20
115, 116 , 117
General specifications
Maximum voltage
(between any terminal and earth ground)
600 V rms
Safety IEC 61010-1, Pollution Degree 2
IEC 61010-2-033 CAT III 600 V, 10 A
EMC IEC 61326-1: Portable
Fuse for A input (115, 117 only) 11 A, 1000 V, IR 17 kA (Fluke PN 803293)
Display Digital: 6,000 counts, updates 4/sec; Bar Graph: 33 segments, updates 32/sec
Temperature
Operating -10 °C to +50 °C
Storage -40 °C to +60 °C
Humidity 0 % to 90 % to 35 °C; 75 % to 40 °C; 45 % to 50 °C
Temperature coefficient 0.1 x (specified accuracy/°C) (< 18 °C or > 28 °C)
Operating altitude 2,000 meters
Battery 9 Volt Alkaline (IEC 6LR61)
Battery life 400 hours typical, without backlight
Certifications CE, CSA, RCM
IP rating (dust and water protection) IP42
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity of
0 % to 90 %.
1 All ac ranges except Auto-V LoZ are specified from 1 % to 100 % of range. Auto-V LoZ is specified from 0.0 V. Because inputs below 1 % of range are not specified, it is
normal for this and other true-rms meters to display non-zero readings when the test leads are disconnected from a circuit or are shorted together. For volts, crest factor of
≤ 3 at 4000 counts, decreasing linearly to 1.5 at full scale. For amps, crest factor of ≤ 3. AC volts is ac-coupled. Auto-V LoZ, ac mV, and ac amps are dc-coupled.