User's Manual
True RMS Multimeters
Testing the Fuses
7
Testing the Fuses
XWWarning
To avoid electrical shock or personal injury:
• Remove the test leads and any input signals before replacing the battery or
fuses.
• To avoid electrical shock, arc blast, or damage to the Meter, install only
fuses with the amperage, interrupt, voltage, and speed ratings specified in
Table 5.
To test the fuses (refer to Figure 1):
1. Set the Rotary Switch to e.
2. Plug a test lead into the VeG terminal and touch the probe to the
400 mA (to test the 440 mA Fuse)
or
10 A terminal (to test the 11 A Fuse).
• If the display shows a resistance value in the range shown in Figure 1, the fuse is good.
• If the display reads
OL, replace the fuse and test again.
• If the display shows any other value, have the Meter serviced. See “Service Information” earlier in
this manual.
11 A 440 mA
Ω
<.5
Ω
<12
OKOK
OK
OK
RANGEHOLD
MIN MAX
V
aik12f.eps
Figure 1. Testing the Current Fuses