User's Manual

175, 177, & 179
Calibration Information
18
Table 4. Calibration Steps (continued)
If the Meter has a serial number from 83911001 to 83911240 or 84060001 and above, or if the model number
flashes on the display at turn on, use the following steps.*
Function
(Switch Position)
Calibration Step Input Value
C-01 600.0 mV dc
E
(DC Millivolts)
C-02
120.0 mV dc
C-03 6.000 V dc
C-04 60.00 V dc
L
(DC Volts)
C-05 600.0 V dc
C-06 600.0 mV, 60 Hz
C
(AC Volts)
C-07 600.0 V, 60 Hz
C-08
600.0 Ω
C-09 6.000 kΩ
C-10 60.00 kΩ
C-11 600.0 kΩ
e
(Ohms)
C-12 6.000 MΩ
O
(Diode Test)
C-13 5.000 V dc
C-14 400.0 mA dc mA
(Milliamps)
C-15 400.0 mA ac, 60 Hz
C-16 6.000 A dc A
(Amps)
C-17 6.000 A ac, 60 Hz
* If the Meter is not connected correctly, or if the rotary switch is in the wrong position, the Meter will beep 2 times to alert the user.