User's Manual

True RMS Multimeters
Calibration Adjustment
17
Calibration Seal
Calibration Button
aik14f.eps
Figure 4. Calibration Access
Table 4. Calibration Steps
If the Meter has a serial number of 84060000 or below, or 83911241 to 84060000, use the following steps.
Function
(Switch Position)
Calibration Step Input Value
C-01 6.000 V, 900.0 Hz
C-02
600.0 mV, 100 Hz
C
(AC Volts)
C-03 6.000 V, 100 Hz
C-04 6.000 V dc
C-05 60.00 V dc
C-06 600.0 V dc
L
(DC Volts)
C-07 1000 V dc
C-08
600.0 mV dc
T E
DC Millivolts
(temperature - 179 only)*
C-09 0.0 °C
e
(Ohms)
C-09 (175/177)
C-10 (179)
6.000 kΩ
R
(Continuity)
C-10 (175/177)
C-11 (179)
600 Ω
mA
(Milliamps)
C-11 (175/177)
C-12 (179)
600.0 mA dc
A
(Amps)
C-12 (175/177)
C-13 (179)
6.000 A dc
* To ensure accuracy, the Meter and Thermocouple Adapter must be at the same temperature. Allow 20 minutes before performing the
temperature procedures
.