User's Manual
True RMS Multimeters
Performance Tests
13
Table 2. Performance Tests 175 (continued)
Meter Response
Test
(Switch
Position)
5500 Output
Lower Limit Upper Limit
5 V 4.990 V dc 5.010 V dc
300 V 299.3 V dc 300.7 V dc
1000 V 996 V dc 1004 V dc
L
DC Volts
-1000 V -1004 V dc -996 V dc
3 V 45 Hz 44.94 Hz 45.06 Hz D Hz
DC Volts
Frequency
1
30 V 50 kHz 49.94 kHz 50.06 kHz
30 mV 0 Hz 29.8 mV dc 30.2 mV dc
-300 mV -300.7 mV dc -299.3 mV dc
E
DC Millivolts
600 mV 598.9 mV dc 601.1 mV dc
500 Ω 495.3 Ω 504.7 Ω e
Ohms
2
19 MΩ 18.68 MΩ 19.32 MΩ
E
Capacitance
1
Apply 0.90 μF 887 nF 913 nF
25 Ω Beeper On R
Continuity
250 Ω Beeper Off
G
Diode
1
Apply 2.0 V 1.978 V dc 2.022 V dc
3 mA 45 Hz 2.92 mA ac 3.08 mA ac
50 mA 1 kHz 49.22 mA ac 50.78 mA ac
F
AC Milliamps
400 mA 1 kHz 393.7 mA ac 406.3 mA ac
4.0 A 45 Hz 3.937 A ac 4.063 A ac ?
AC Amps
9.0 A 1 kHz 8.83 A ac 9.17 A ac
3 mA, 0 Hz 2.94 mA dc 3.06 mA dc
50 mA 49.47 mA dc 50.53 mA dc
mI
DC Milliamp
1
-400 mA -404.3 mA dc -395.7 mA dc
4.0 A 3.957 A dc 4.043 A dc A
DC Amps
1
-9.0 A -9.12 A dc -8.88 A dc
1. Press the YELLOW button to access this function.
2. Does not include test lead resistance.