User's Manual
True RMS Multimeters
Performance Tests
11
Required Equipment
A Fluke 5500A Multi-Product Calibrator (or equivalent) is required for the performance test procedures in
this document.
If an equivalent calibrator is used, it must meet the accuracy specifications shown in Table 1.
Table 1. Calibrator Specifications
Recommended Equipment Measurement Function Accuracy
5500A Multi-Product Calibrator
(or equivalent)
DC Volts 30 mV to 1000 V
±0.0225 %
DC Current 3 mA to 9 A
±0.25 %
AC Volts 300 mV to 1000 V
±0.25 % @ 45 Hz to 1 kHz
AC Current 3 mA to 9 A
±0.375 % @ 45 Hz to 1 kHz
Resistance 25 Ω to 500 Ω
±0.225 %
19 M Ω
±0.4 %
Capacitance 900 nF
±0.30 %
Temperature -40 to 400 °C
0.25 %
Frequency 5 V, 50 kHz
±0.025 %
Fluke 80 AK
K-type Thermocouple Adapter Accessory
Temperature -
K-type Thermocouple, mini-plug on both
ends
Temperature -
Preparing for the Performance Test
XWWarning
To avoid possible electric shock or personal injury:
• Do not perform the following procedures unless qualified to do so. Some
procedures involve the use of high voltages.
• Before handling the test connections and in between tests, make sure the
calibrator is in standby mode (STBY).
To prepare for the performance test:
1. Make sure that you have the required equipment (refer to Table 1).
2. Warm up the calibrator as required by its specifications.
3. Allow the temperature of the UUT to stabilize at room temperature (23 °C ± 5 °C [73 °F ± 9 °F]).
4. Check the fuses and Battery, and replace them if necessary. Refer to "Testing the Fuses”, “Replacing
the Fuses", and "Replacing the Battery".










