Detailed Specifications
Accuracy ± ( [ % of Reading ] + [ Counts ] )
Function Range
1
Resolution Model 175 Model 177 Model 179
AC Volts
2
600.0 mV 0.1 mV 1.0 % + 3 1.0 % + 3 1.0 % + 3
6.000V 0.001V (45 Hz to 500 Hz) (45 Hz to 500 Hz) (45 Hz to 500 Hz)
60.00V 0.01V
600.0V 0.1V
1000V 1V 2.0 % + 3 2.0 % + 3 2.0 % + 3
(500 Hz to 1 kHz) (500 Hz to 1 kHz) (500 Hz to 1 kHz)
DC mV 600.0 mV 0.1 mV 0.15 % + 2 0.09 % + 2 0.09 % + 2
DC Volts 6.000V 0.001V 0.15 % + 2 0.09 % + 2 0.09 % + 2
60.00V 0.01V
600.0V 0.01V
1000V 1V 0.15 % + 2 0.1 % + 2 0.1 % + 2
Continuity 600 1 Meter beeps at < 25 , beeper turns off at > 250 ;
detects opens or shorts of 250 ms or longer.
Ohms 600.0 0.1 0.9 % + 2 0.9 % + 2 0.9 % + 2
6.000 k 0.001 k 0.9 % + 1 0.9 % + 1 0.9 % + 1
60.00 k 0.01 k 0.9 % + 1 0.9 % + 1 0.9 % + 1
600.0 k 0.1 k 0.9 % + 1 0.9 % + 1 0.9 % + 1
6.000 M 0.001 M 0.9 % + 1 0.9 % + 1 0.9 % + 1
50.00 M 0.01 M 1.5 % + 3 1.5 % + 3 1.5 % + 3
Diode test 2.400V 0.001V 1 % + 2
Capacitance 1000 nF 1 nF 1.2 % + 2 1.2 % + 2 1.2 % + 2
10.00 µF0.01 µF 1.2 % + 2 1.2 % + 2 1.2 % + 2
100.0 µF0.1 µF 1.2 % + 2 1.2 % + 2 1.2 % + 2
9999 µF
3
1 µF 10 % typical 10 % typical 10 % typical
AC Amps 60.00 mA 0.01 mA 1.5 % + 3 1.5 % + 3 1.5 % + 3
(True-rms) 400.0 mA 0.1 mA
(600 mA for 18 hrs) 0.001A
6.000A
(45 Hz to 1 kHz) 10.00A 0.01A
(20A for 30s)
DC Amps 60.00 mA 0.01 mA 1.0 % + 3 1.0 % + 3 1.0 % + 3
400.0 mA 0.1 mA
(600 mA for 18 hrs)
6.000A 0.001A
10.00A 0.01A
(20A for 30s)
Hz 99.99 Hz 0.01 Hz 0.1 % + 1 0.1 % + 1 0.1 % + 1
(AC- or DC- 999.9 Hz 0.1 Hz
coupled, 9.999 kHz 0.001 kHz
V or A
4, 5
input ) 99.99 kHz 0.01 kHz
Temperature -40 °C to +400 °C0.1 °C NA NA 1 % + 1.0 °C
-40 °F to +752 °F0.1 °F 1 % + 1.8 °F
MIN MAX AVG For DC functions, accuracy is the specified of the measurement function ± 12 counts for changes
longer than 275 ms in duration.
For AC functions, accuracy is the specified of the measurement function ± 40 counts for changes
longer than 1.2 s in duration.
1. All AC voltage and AC current ranges are specified from 5 % of range to 100 % of range.
2. Crest factor of 3 at full scale up to 500 V, decreasing linearly to crest factor 1.5 at 1000 V.
3. In the 9999 µF range for measurements to 1000 µF, the measurement accuracy is 1.2 % for all models.
4. In mA and A ranges, frequency measurement is specified to 30 kHz.
5. Frequency < 10 kHz are not specified in 600 mV AC, 60 mA AC, and 6 A AC ranges.
Function Overload Protection
1
Input Impedance Common Mode Rejection Ratio Normal Mode Rejection
(Nominal) (1 k Unbalanced)
Volts AC 1000 V RMS or DC > 10 M < 100 pF > 60 dB @ DC, 50 or 60 Hz
Volts DC 1000 V RMS or DC > 10 M <100 pF >120 dB @ DC, 50 or 60 Hz > 60 dB @ 50 Hz or 60 Hz
Open Circuit Test Full Scale Voltage To: Short Circuit Current
Voltage 6.0 M 50 M
Ohms 1000V RMS or DC < 1.5 V DC < 600 mV DC < 1.5 V DC < 500 µA
Diode test 1000V RMS or DC 2.4 to 3.0 V DC 2.4 V DC < 1.2 mA typical
1. 10
7
V-Hz maximum.