Data Sheet
Page 2 of 3, Document #2155
170 Series True-rms DMMs
Extended Specifications
©2000 Fluke Corporation
Rev. A-11/2000
General Specifications
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, with relative
humidity of 0 % to 75 %. Accuracy specifications take the form of: ±  ( [ % of Reading ] + [ Counts ] )
Maximum voltage between any 1000V DC or AC RMS
terminal and earth ground
Surge Protection 8 kV peak per IEC 61010
Fuse for mA inputs 440 mA, 1000 V FAST Fuse
Fuse for A input 11A, 1000V FAST Fuse
Display Digital: 6,000 counts, updates 4/sec
Bar Graph: 33 segments, updates 40/sec
Frequency: 9,999 counts
Capacitance: 9,999 counts
Altitude Operating: 2000 m; Storage: 12000 m
Temperature Operating: -10 °C to +50 °C
Storage: -30 °C to +60 °C
Temperature coefficient 0.1 X (specified accuracy / °C)
(< 18 °C or > 28 °C)
Electromagnetic Compatibility In an RF field of 3 V/M, accuracy = specified accuracy except in
(EN 61326-1:1997) temperature: specified accuracy ± 5 °C, ± 9 °F
Relative Humidity 0 % to 90 % @ 0 °C to 35 °C;
0 % to 70 % @ 36 °C to 50 °C
Relative Humidity in 0 % to 80 % @ 0°C to 35°C;
50 MΩ Range 0 % to 70 % @ 36 °C to 50 °C
Battery Life Alkaline: ~200 hrs typical
Size, with Holster (H x W x L) 4.3 cm x 9 cm x 18.5 cm
Weight 420g
Safety Compliances ANSI/ISA S82.02.01, CSA C22.2-1010.1, IEC 61010 to 1000 V Overvoltage
Category III, 600 V Overvoltage Category IV
Certifications CSA, TÜV (EN61010),
Australian (N10140)